Appendix A
SCPI Command Reference
MEASure subsystem
4460 GSM System Option and 4468 EDGE System Option
Version 12.20
495
:MEASure:GSM:ARRay:RFRX:BER:ALL
Description
Performs the measurement of the frame erasure ratio a specific number of times (set
with the
int1
parameter). To read out
the entire
measurement results array, use
the
:FETCh:GSM:RFRX:RBER:FER
command.
Notes:
–
More details regarding RBER testing can be found in section BER Live Display.
–
The number of samples to be used for testing can be set, using the
:CONF:GSM:BER:COUNt
command.
Query
The query form of this command will perform the measurement the specified number
of times (
int1
parameter). As soon as all measurements have been completed, all
measurement result values will be returned in a string. The string delivered back will
contain
int1
floating point real numbers, representing the frame erasure ratios.
The single measurement results are separated by commas.
Example
:MEASure:GSM:ARRay:RFRX:RBER:FER 5
:FETCh:GSM:RFRX:RBER:FER?
The string returned is:
"0.0,0.1,0.0,0.2,0.1"
.
Syntax
:MEASure:GSM:ARRay:RFRX:BER:ALL <int1>
Parameters
int1
is an integer.
The minimum value for
int1
is
0
, the maximum value is
100
. The default value for
int1
is
0
.
Description
Performs a sequential measurement of the (nonresidual) residual bit error ratios
(RBER) a specific number of times (set with the
int1
parameter). To read out
the
entire
measurement results array, use the
:FETCh:GSM:RFRX:BER:ALL
com-
mand.
The RF RX tests performed by this command are:
CIA
, the (nonresidual) bit error ratio of the class Ia bits,
CIB
, the (nonresidual) bit error ratio of the class Ib bits, and
CII
, the (nonresidual) bit error ratio of the class II bits.
Notes:
–
More details regarding BER testing can be found in section BER Live Display.
–
More details regarding bit classes can be found in section Voice Coding.
–
Please keep in mind that the start of a new RF RX test will always terminate the
preceding one instantly.
–
The number of samples to be used for testing can be set, using the
:CONF:GSM:BER:COUNt
command.
–
For further details regarding group measurements, please refer to section
“:MEA-
Sure:...:GROup” on page 360
.
–
More information regarding
ARRay
measurements can be found in section
“MEASure:ARRay” on page 356
.
Query
The query form of this command will perform the measurements the specified num-
ber of times (
int1
parameter). After all measurements have been completed and all
measurement results obtained, the query delivers a string, containing (3 *
int1
)
measurement result values. The single measurement result values are separated by
commas. All measurement result values are floating point real numbers. The order of
the measurement result values delivered back is as follows:
1.
CIA
, representing the (nonresidual) bit error ratio of the class Ia bits,
2.
CIB
, representing the (nonresidual) bit error ratio of the class Ib bits, and
3.
CII
, representing the (nonresidual) bit error ratio of the class II bits.