Appendix A
SCPI Command Reference
CALCulate Subsystem
700
4460 GSM System Option and 4468 EDGE System Option
Version 12.20
:CALCulate:EGPRs:RFSPectrum:ACPM:MODulation:LIMit
[:FAIL]?
:CALCulate:EGPRs:RFSPectrum:ACPM:MODulation
:LIMit:STATe
Query
Returns the
mean
value and the
standard deviation
of the RF spectrum measure-
ment.
In the case of the ACP Mod measurement, the string delivered back contains 27 pairs
of mean values and standard deviation values for the ACP Mod measurement.
In the ACP Transient case, 23 pairs of mean values and standard deviation values are
returned.
The values are floating point real values and separated by commas.
Example
:CALC:RES
:MEAS:EGPR:ARR:RFSP:ACPM:MOD
...(other SCPI commands)...
:CALC:EGPR:RFSP:ACPM:MSIG?
Syntax
:CALCulate:EGPRs:RFSPectrum:ACPM
:MODulation:LIMit[:FAIL]?
Parameters
There are no parameters.
Description
There is solely a query form of this command available.
Query
This command delivers 27 boolean numbers in a string, separated by commas. These
numbers indicate whether the 27 measurement result values of the
MEAS:EGPRs:RFSP:ACPM:TRAN
measurement did violate their individual lim-
its (set with the commands explained below). Every boolean number represents the
result of the limit check for one of the 27 measurement result values. The first value
will give the result of the limit check at -1.83 MHz offset to the carrier, the second
one for -1.80 MHz offset to the carrier and so on.
While a
0
delivered on any position means that the related measurement result value
did not exceed the limits, a
1
indicates that at least one limit was violated by the
measurement result value.
Note According to GSM specification 05.05, this measurement will be performed on
0, (±)400, (±)600, (±)1.200 and (±)1.800 kHz offset from the carrier. All measure-
ments will be performed with a measuring bandwidth of 30 kHz. For additional infor-
mation, also the measurement values of the 30 kHz windows below and above the
frequencies specified by GSM standard 05.05 will be taken into account. This adds up
to 9 x 3 = 27 measurement result values.
Example
:MEAS:EGPR:BLOC:MSP:AVG 20
:CALC:EGPR:RFSP:ACPM:MOD:LIM?
String returned in this example: "
0,0,0,0,0,0,0,0,1,
0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0
"
The
1
on position 9 indicates that there was a violation of at least one limit by the
measurement result value taken at an offset of -0.57 MHz from the carrier.
Syntax
:CALCulate:EGPRs:RFSPectrum:ACPM
:MODulation:LIMit:STATe <PredefExp>
Parameters
PredefExp
is one of the following predefined expressions:
ON|OFF
. Default is
ON
.