Appendix A
SCPI Command Reference
RFGenerator subsystem
748
4460 GSM System Option and 4468 EDGE System Option
Version 12.20
:RFGenerator:GSM:STATe
:RFGenerator:GSM:LEVel
:RFGenerator:GSM:FREQuency
Description
This command switches between burst transmission (
BURSt
) or continuous signal
transmission without bursts (
CWAVe
).
CWAVe
means that Willtek 4400 will generate a coded GSM signal in all timeslots
according to the settings above.
When set to
BURSt
, Willtek 4400 will generate only one burst per TDMA frame. This
means that there will be no RF transmission during the remaining 7 timeslots.
Query
The query form of this command returns the current setting. A query will return a
string, containing the short-form version of one of the predefined expressions
explained above.
Example
:RFGenerator:EGPRs:STYPe BURSt
:RFG:EGPRs:STYP?
Value returned in this example "
BURSt
".
Syntax
:RFGenerator:GSM:STATe <PredefExp>
Parameters
PredefExp
is one of the following predefined expressions:
ON|OFF
.
Default is
OFF
.
Description
Switches the RF generator on or off.
While the Willtek 4400 is in RF generator mode, there is no signaling active. This
means that the Willtek 4400 may be used as RF signal generator for all kinds of RF
and GSM signals. While working as an RF generator, the Willtek 4400 does not
respond to any messages sent by the mobile nor does the test set expect the mobile
under test to react in any way.
Query
The query form of this command returns the current setting. A query will return a
string, containing one of the predefined expressions explained above.
Example
:RFGenerator:GSM:STATe ON
:RFG:GSM:STAT?
Value returned:
"ON"
.
Syntax
:RFGenerator:GSM:LEVel <real1>
Parameters
real1
is a floating point real number. The minimum value for
real1
is
-120.0
,
the maximum value is
-10.0
. The minimum resolution for
real1
is
0.1
. The
default value for
real1
is
-60.0
.
Description
This command sets the RF output power level of the Willtek 4400. The value specified
for
real1
is the power output level in dBm.
Query
The query form of this command will return the current setting. The string delivered
back will contain one floating point real number.
Example
:RFGenerator:GSM:LEVel -50.5
:RFG:GSM:LEV?
Value returned:
"-50.5"
.
Syntax
:RFGenerator:GSM:FREQuency <real1>