Appendix A
SCPI Command Reference
MEASure subsystem
500
4460 GSM System Option and 4468 EDGE System Option
Version 12.20
:MEAS:GSM:ARR:RFSP:ACPM:TRANsient
:MEASure:ARRay:AFANalyser:GROup
Query
The query form of this command will perform the measurement the specified number
of times (
int1
parameter). As soon as all measurements have been completed, all
measurement result values will be returned in a string. The string delivered back will
contain the absolute power at the center frequency (within a 30 kHz bandwidth) as
the first value with the physical dimension of dBm and after that 23 floating point
real numbers with the physical dimension dBc. The total count of the values is 24 *
int1
.
The individual measurement results are separated by commas.
Example
:MEAS:ARR:RFSP:ACPM:MODP 5
:FETC:RFSP:ACPM:MODP?
The string returned in this example would contain 120 floating point real numbers,
representing the 24 measurement result values of five measurement runs.
Syntax
:MEAS:GSM:ARR:RFSP:ACPM:TRANsient <int1>
Parameters
int1
is an integer.
The minimum value for
int1
is
0
, the maximum value is
100
. The default value for
int1
is
0
.
Description
Performs the measurement of the ACPM spectrum due to switching transients a spe-
cific number of times (set with the
int1
parameter). To read out
the entire
mea-
surement results array, use the
:FETCh:GSM:RFSP:ACPM:TRAN
command.
Note:
According to GSM specification 05.05, this measurement will be performed on
0, (±)400, (±)600, (±)1.200 and (±)1.800 kHz offset from the carrier. All measure-
ments will be performed with a measuring bandwidth of 30 kHz. For additional
information, also the measurement values of the adjacent 30 kHz windows below
and above the frequencies given above will be delivered back. This adds up to 9 x 3 =
27 measurement result values, that are returned per measurement run.
More information regarding
ARRay
measurements can be found in section
“MEA-
Sure:ARRay” on page 356
.
Query
The query form of this command will perform the measurement the specified number
of times (
int1
parameter). As soon as all measurements have been completed, all
measurement result values will be returned in a string. The string delivered back will
contain (27 *
int1
) floating point real numbers. The physical dimension is dBm.
The single measurement results are separated by commas.
Example
:MEAS:GSM:ARR:RFSPectrum:ACPM:TRANsient 5
:FETCh:GSM:RFSP:ACPM:TRAN?
The string returned would contain 135 floating point real numbers, representing the
27 measurement result values of five measurement runs.
Syntax
:MEASure:ARRay:AFANalyser:GROup <int1>
Parameters
int1
is an integer.
The minimum value for
int1
is
0
, the maximum value is
100
. The default value for
int1
is
0
.