Appendix A
SCPI Command Reference
CALCulate Subsystem
4460 GSM System Option and 4468 EDGE System Option
Version 12.20
623
:CALCulate:GSM:RFTX:POWer:LIMit:STATe
:CALCulate:GSM:RFTX:POWer:LIMit:UPPer[:DATA]
:CALCulate:GSM:RFTX:POWer:LIMit:LOWer[:DATA]
Query
This command delivers one boolean number, indicating whether the user-definable
limits of the measurement of the mobile's RF output power level were violated by a
measurement result value or not.
A
0
means that all measurement result values were within the limits set, while a
1
indicates that at least one measurement result value did violate at least one of the
limits.
The limits can be set using the commands described below.
Example
:MEAS:GSM:ARRay:POW 20
:CALC:GSM:RFTX:POW:LIM?
String returned:
"0"
This string indicates that there was no violation of the measurement limits.
Syntax
:CALCulate:GSM:RFTX:POWer:LIMit:STATe <PredefExp>
Parameters
PredefExp
is one of the following predefined expressions:
ON|OFF
.
Default is
ON
.
Description
This command switches the check of the measurement result values of the
MEAS:GSM:RFTX:POWer
measurement against their user-definable limits either
on or off.
Query
There is no query form of this command available.
Example
:CALC:GSM:RFTX:POW:LIM:STAT ON
Switches the limit check for the RF output power level measurement on.
Syntax
:CALCulate:GSM:RFTX:POWer:LIMit:UPPer[:DATA] <real1>
Parameters
real1
is a floating point real number. The minimum value is
-120.0
, the maxi-
mum is
50.0
. The minimum resolution possible is
0.1
, the default value is
39
.
Description
Sets the
upper
limit for the analyzer's RF input power level. The physical dimension of
the number stated is dBm.
Query
There is no query form of this command available.
Example
:CALC:GSM:RFTX:POW:LIM:UPP 13
Sets the upper limit of the peak power measurement to 13 dBm.
Syntax
:CALCulate:GSM:RFTX:POWer:LIMit:LOWer[:DATA] <real1>
Parameters
real1
is a floating point real number. The minimum value is
-120.0
, the maxi-
mum is
50.0
. The minimum resolution possible is
0.1
, the default value is
-60.0
.
Description
Sets the
lower
limit for the analyzer's RF input power level. The physical dimension of
the number stated is dBm.
Query
There is no query form of this command available.