Appendix A
SCPI Command Reference
RFGenerator subsystem
758
4460 GSM System Option and 4468 EDGE System Option
Version 12.20
:RFGenerator:EGPRs:MODulation:TSEQuence:[DATA]
:RFGenerator:EGPRs:MSLOT:LEVel:[DATA]
Description
This command switches the use of a training sequence either ON or OFF.
Notes:
–
The type of training sequence can be set using the command described below.
–
More details regarding the training sequence can be found in section Timeslots
and Bursts.
Query
The query form of this command returns the current setting. A query will return a
string, containing one of the predefined expressions explained above.
Example
:RFG:EGPRs:MODulation:TSEQuence:STATe ON
:RFG:EGPRs:MOD:TSEQ:STAT?
Value returned in this example "
ON
".
Syntax
:RFGenerator:EGPRs:MODulation:TSEQuence:[DATA] <int1>
Parameters
int1
is an integer.
The minimum value for
int1
is
0
, the maximum value is
7
. The default value is
0
.
Note:
This command only selects the type of training sequence. To switch the use of
a training sequence on or off, use the command described above.
Description
This command selects the type of training sequence to be used.
More details regarding the training sequence can be found in section Timeslots and
Bursts.
Query
The query form of this command returns the current setting. A query will return a
string, containing one integer.
Example
:RFG:EGPRs:MODulation:TSEQuence:DATA 3
:RFG:EGPRs:MOD:TSEQ?
Value returned in this example "
3
".
Syntax
:RFGenerator:EGPRs:MSLOT:LEVel:[DATA] <real1>
Parameters
real1...real8
are eight floating point real numbers.
The minimum value for each
realx
is
-120.0
, the maximum value is
-10.0
. The
resolution of each
realx
is
0.1
. The default value for each
realx
is
-60.0
.
Note:
All eight values must be specified.
Description
This command configures individual RF output power levels for the eight DL timeslots.
Every value represents the RF output power level of Willtek 4400 in dBm during the
related timeslot.
Note:
This command
only prepares
the use of individual RF power levels. To activate
them, use the
RFG:MSL:LEV:MODE
command.
Query
The query form of this command will return the current setting. Always eight floating
point real numbers will be returned in a string.
Example
:RFGenerator:EGPRs:MSLot:LEVel:DATA -80.0,-70.0,-60.0,-
50.0,-40.0, -30.0,-20.0,-10.0
:RFG:MSL:LEV?
Values returned in this example: "
-80.0,-70.0,-60.0,-50.0,-40.0,
-30.0,-20.0,-10.0
"