:MEASure:EGPRs[:CONTinuous]:RFTX:ERMS
:MEASure:EGPRs[:CONTinuous]:RFTX:EPEAk
Example
:MEASure:EGPRs:CONTinuous:RFTX:ALL?
In this case, all relevant EGPRS RF TX measurements will be performed in a sequence.
As soon as all measurements have been completed and all measurement results
obtained, a string will be delivered back containing 17 measurement results:
"
5.13,11.94,2.43,4.55,-2.22,557.0,0.1,11.22,0,-72.18, -
61.91,-20.91,-0.05,-0.04,-17.97,-56.60, -73.95
".
Syntax
:MEASure:EGPRs[:CONTinuous]:RFTX:ERMS
Parameters
There are no parameters.
Description
Starts a continuous measurement of the error vector magnitude rms-averaged over a
burst. EVM is a relative value and shown as a percentage. To read out the measure-
ment results, use the
:FETCh:EGPRs:RFTX:ERMS
command.
Query
The query form of this command will start the measurement and return a measure-
ment result value as soon as the first measurement has been completed. The string
delivered back will contain one floating point real number with the physical dimen-
sion of %.
Example
:MEAS:EGPRs:ARR:RFTX:ERMS 5
:FETC:EGPRs:RFTX:ERMS?
The error vector magnitude rms-averaged measurement will be performed five times.
The measurement results will be stored internally and read by the <TT>FETC</TT>
command. An example of the string returned is:
"
2.33,2.53,1.97,2.41,1.79
".
Syntax
:MEASure:EGPRs[:CONTinuous]:RFTX:EPEAk
Parameters
There are no parameters.
Description
Starts a continuous measurement of the maximum error vector magnitude within a
burst. EVM is indicated per cent. To read out the measurement results, use the
:FETCh:EGPRs:RFTX:EPEAk
command.
Query
The query form of this command will start the measurement and return a measure-
ment result value as soon as the first measurement has been completed. The string
delivered back will contain one floating point real number with the physical dimen-
sion of %.
Example
:MEAS:EGPRs:ARR:RFTX:EPEAk 5
:FETC:EGPRs:RFTX:EPEAk?
The maximum error vector magnitude measurement will be performed five times. The
measurement results will be stored internally and read by the
FETC
command. An
example of the string returned is: "