Appendix A
SCPI Command Reference
CALCulate Subsystem
4460 GSM System Option and 4468 EDGE System Option
Version 12.20
665
:CALCulate:AFANalyser:FREQuency:LIMit:STAte?
:CALCulate:AFANalyser
:FREQuency:LIMit:UPPer[:DATA]
:CALCulate:AFANalyser
:FREQuency:LIMit:LOWer[:DATA]
:CALCulate:AFANalyser
:DISTortion:LIMit[:FAIL]?
Syntax
:CALCulate:AFANalyser:FREQuency:LIMit:STAte?
<PredefExp>
Parameters
PredefExp
is one of the following predefined expressions:
ON|OFF
. Default is
ON
.
Description
This command switches the check of the measurement result values of the
MEAS:AFAN:FREQ
measurement against their user-definable limits either on or
off.
Query
There is no query form of this command available.
Example
:CALC:AFAN:FREQ:LIM:STAT ON
Switches the limit check of the AF frequency measurement on.
Syntax
:CALCulate:AFANalyser
:FREQuency:LIMit:UPPer[:DATA] <real1>
Parameters
real1
is a floating point real number. The minimum value is
0.0
, the maximum is
20,000.0
. The minimum resolution possible is
1.0
, the default value is
5,000.0
.
Description
Sets the
upper
limit for the AF frequency measurement. The physical dimension of the
number stated is Hertz.
Query
The query form of this command is not available.
Example
:CALC:AFAN:FREQ:LIM:UPP 10000
Sets the upper limit of the audio frequency measurement to 10 kHz.
Syntax
:CALCulate:AFANalyser
:FREQuency:LIMit:LOWer[:DATA] <real1>
Parameters
real1
is a floating point real number. The minimum value is
0.0
, the maximum is
20,000.0
. The minimum resolution possible is
1.0
, the default value is
1,000.0
.
Description
Sets the
lower
limit for the AF frequency measurement. The physical dimension of the
number stated is Hertz.
Query
The query form of this command is not available.
Example
:CALC:AFAN:FREQ:LIM:LOW 950
Sets the lower limit of the audio frequency measurement to 950 Hz.
Syntax
:CALCulate:AFANalyser
:DISTortion:LIMit[:FAIL]?