Appendix A
SCPI Command Reference
MEASure subsystem
4460 GSM System Option and 4468 EDGE System Option
Version 12.20
479
:MEASure[:CONTinuous]:AFANalyser:STOP
:MEASure[:CONTinuous]:AFANalyser:GROup
Description
Starts a continuous measurement of the ACPM spectrum due to switching tran-
sients. To read out the latest measurement results, use the
:FETCh:GSM:RFSP:ACPM:TRANS
command.
Note:
According to GSM specification 05.05, this measurement will be performed on
0, (±)400, (±)600, (±)1.200 and (±)1.800 kHz offset from the carrier. All measure-
ments will be performed with a measuring bandwidth of 30 kHz. For additional
information, also the measurement values of the 30 kHz windows below and above
the frequencies specified by GSM standard 05.05 will be delivered back. This adds up
to 9 x 3 = 27 measurement result values, that are returned.
Query
The query form of this command starts the measurement and delivers a string, con-
taining 27 measurement result values separated by commas. These 27 measurement
result values are floating point real numbers with the physical dimension dBm.
Example
:MEAS:GSM:CONT:RFSP:ACPM:TRANsient?
In this case, a measurement of the modulation spectrum will be performed. As soon
as the measurement has been completed, a string will be delivered back containing
27 floating point real numbers.
Syntax
:MEASure[:CONTinuous]:AFANalyser:STOP
Parameters
There are no parameters.
Description
Stops continuous audio measurements.
Note
Stopping the test is not required since a test is automatically stopped when you
start a new test.
Query
The query form of this command is not available.
Example
:MEAS:AFAN:SIN
:MEAS:AFAN:STOP
Starts and stops a measurement.
Syntax
:MEASure[:CONTinuous]:AFANalyser:GROup
Parameters
There are no parameters.
Description
Starts a continuous measurement of the audio tests, specified with the
:CONF:MEAS:GRO:AFAN
command. To read out the latest measurement results,
use the
:FETCh:AFAN:GROup
command.
Notes:
–
Please keep in mind that the start of a new AF test will always terminate all
other measurements (see section
“The MEASure subsystem” on page 353
for
details).
–
To perform any audio measurements on your Willtek 4400, the Audio Option
must be installed.
–
For further details regarding group measurements, please refer to section
“:MEA-
Sure:...:GROup” on page 360
.