Appendix A
SCPI Command Reference
MEASure subsystem
504
4460 GSM System Option and 4468 EDGE System Option
Version 12.20
:MEASure:ARRay:AFAN:FREQuency
:MEASure:ARRay:AFAN:DISTortion
:MEASure:ARRay:AFANalyser:SINad
Example
:MEAS:ARR:AFANanlyser:DCVoltage? 5
The string returned is:
"0.025,0.029,0.019,0.030,0.025"
.
Syntax
:MEASure:ARRay:AFAN:FREQuency <int1>
Parameters
int1
is an integer.
The minimum value for
int1
is
0
, the maximum value is
100
. The default value for
int1
is
0
.
Description
Performs the measurement of the frequency of the audio signal applied to the audio
analyzer a specific number of times (set with the
int1
parameter). To read out
the
entire
measurement results array, use the
:FETCh:AFAN:FREQ
command.
Note:
Any AF test command will need the Audio Option to be installed on your
Willtek 4400.
Query
The query form of this command will perform the measurement the specified number
of times (
int1
parameter). As soon as all measurements have been completed, all
measurement result values will be returned in a string. The string delivered back will
contain
int1
floating point real numbers. The physical dimension of the measure-
ment result values is Hertz.
The single measurement results are separated by commas.
Example
:MEASure:ARRay:AFANanlyser:FREQ? 5
The string returned is:
"1000.0,1000.5,1000.9,999.9,1000.0"
.
Syntax
:MEASure:ARRay:AFAN:DISTortion <int1>
Parameters
int1
is an integer.
The minimum value for
int1
is
0
, the maximum value is
100
. The default value for
int1
is
0
.
Description
Performs the measurement of the distortion of a sine wave applied to the audio ana-
lyzer a specific number of times (set with the
int1
parameter). To read out
the
entire
measurement results array, use the
:FETCh:AFAN:DIST
command.
Note:
Any AF test command will need the Audio Option to be installed on your
Willtek 4400.
Query
The query form of this command will perform the measurement the specified number
of times (
int1
parameter). As soon as all measurements have been completed, all
measurement result values will be returned in a string. The string delivered back will
contain
int1
floating point real numbers. The physical dimension of the measure-
ment result values is percentage (%).
The single measurement results are separated by commas.
Example
:MEASure:ARRay:AFANanlyser:DIST? 5
The string returned is:
"0.5,0.7,1.1,1.2,0.9"
.
Syntax
:MEASure:ARRay:AFANalyser:SINad <int1>