Appendix A
SCPI Command Reference
AFANalyser subsystem
802
4460 GSM System Option and 4468 EDGE System Option
Version 12.20
:AFANalyser:SPECtrum:RLEVel
Description
Selects the type of filter to be inserted between the signal input and the AF analyzer.
NONE
means that there will be no filtering.
CCITt
selects a speech filter as defined by CCITT regulations.
CMESsage
is a speech filter according to US standards.
Query
The query form of this command returns the current setting. A query will return a
string, containing the short-form version of one of the predefined expressions
explained above.
Example
:AFANalyser:FILTer NONE
:AFAN:FILT?
Value returned:
"NONE"
.
Syntax
:AFANalyser:SPECtrum:RLEVel <real1>
Parameters
real1
is a floating point real number. The minimum value for
real1
is
0.001
,
the maximum value is
30.0
. The minimum resolution for
real1
is
0.001
. The
default value for
real1
is
1.0
.
Description
This command sets the level of the reference voltage (i.e. the 0 dB line of the AF spec-
trum display). The physical unit of
real1
is volt if the signal source is
AF in
.
Notes
–
When the signal source is
RF in
, the 0 dB line is interpreted as full scale.
–
A typical audio standard is 0.775 volts for 0 dB.
Query
The query form of this command will return the current setting. The string delivered
back will contain one floating point real number.
Example
:AFANalyser:SPECtrum:RLEVel 13.0
:AFAN:SPEC:RLEV?
Value returned:
"13.0"
.