Chapter 3
GSM Asynchronous Mode Operation
Spectrum measurements
4460 GSM System Option and 4468 EDGE System Option
Version 12.20
147
Working with the averaging tool
– Enter the averaging tool by pushing the
|Start Avg.|
softkey.
– The 4400 will now start the averaging process, using the number of
measurements set on the
Avg
entry field.
– As long as the number of measurements taken is less than the number
entered on the
Avg
entry field, the number of the current measurement will
be displayed in red on the
Count
display field.
– As soon as the number of measurements taken is equal or larger than the
number of measurements specified for the averaging process, the number on
the
Count
display field will turn blue.
Important to note
– The entry range for the
Avg
entry field is from 1 to 200.
– The
Avg
entry field is not accessible unless the averaging process has been
started.
– As soon as you change the current setting on the
Avg
entry field, the aver-
aging process will be reset. This means that all previous measurements will
be discharged and the series of measurements will be started anew.
– A reset of the averaging process can also be achieved by pushing the
|Reset|
softkey.
– When the number of measurements taken (display field
Count
) exceeds the
number of measurements specified for the averaging process (
Avg
entry
field), always the last
Avg
measurements will be averaged and displayed in
the live display.
– In case the live display is frozen and unfrozen (using the appropriate softkey)
during an averaging process, the averaging process will not be interrupted
but will continue in the background.
– Even when the live display has been frozen, a reset of the averaging is
possible, pushing the
|Reset|
softkey.
Viewing the maximum deviation
The spectrum live display shows the current or the averaged spectrum of the
signal. In some cases it can be useful to check the maximum deviation from the
current or the average. Push
|Min-Max ON|
for this purpose; the 4400 then
displays two additional curves in red, visualizing the minimum and the maximum
for each frequency point as measured since the start of the test.
Push
|Reset|
to erase previous minima and maxima and to start the measure-
ments all over.
Push
|Min-Max OFF|
to resume measurements without the deviation display.