II-42
f. AGA operation
AGA (Automatic Gain Adjustment) function
The AGA function is performed after the transducers are attached and before measurement
operation in order to set the optimum reception gain.
There are two modes - automatic and manual.
Display Menu No.
: CHK-5: [GAIN]
MANUAL---#A (Gain to be set manually for all paths at once? Yes or No)
#1 (1
st
path only)
|
#4 (4
th
path only)
AUTO------#A (Gain to be set automatically for all paths at once? Yes or No)
#1 (1
st
path only)
|
#4 (4
th
path only)
When you can not complete AGA settings in success and found error message as below, please
refer to below and try again.
Otherwise please contact our representatives.
aa) “bW” : Bubble Warning
The error will be indicated when pulse height fluctuates during AGA setting due to fluid
conditions like air bubble contained.
In such case, you might be required flow condition change or change path method to shorter
path. (ex. V method to Z method.) Then try again AGA function.
bb) “wW” : Wave distortion Warning
The error will be indicated when required ratio of pulse height for AGA setting can not be kept due
to pipe conditions like inner corrosion.
In such case, you might be required mounting position change or change path method to
shorter path. (ex. V method to Z method.) Then try again AGA function.
Summary of Contents for UFL-30
Page 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Page 11: ... 10 blank ...
Page 12: ...I Installation ...
Page 60: ...II Operation ...
Page 109: ...II 50 blank ...
Page 110: ...III Other ...
Page 123: ...III 14 III 2 5 Dimensions Main Unit ...
Page 128: ...III 19 blank ...
Page 129: ...III 20 blank ...
Page 145: ...III 36 blank ...
Page 165: ...III 56 blank ...