I-2-10. Transducer Cable Treatment for Main Unit
Cable treatment will be divided 3 part.
Step1; Outer sheath part
Step2; Inner sheath part
Step3; Core part
Step 1; Outer sheath treatment
1. Measure 40mm from the cable edge.
2. Make slit around cable
circumference
at the
point.
NOTE: Do not damage shield part by over
slitting.
Please remain sheath thickness as
around 0.1mm at least.
3. Make slit to outside direction as below.
4. Remove outer sheath.
5. Fold back outer shield.
40mm
Cable section
Outer sheath
Outer shield
Core
Core sheath
Inner sheath
Inner shield
Cable is composed as below.
I- 36