II-49
- M A X
V A L U E
- - - - - - -
E + 3
m
3
/ h
II-2-8 Error Message
If invalid data or selection is entered, an error message will be displayed. The previous setting is
automatically retained and the display will revert to the previous indication.
ex : When invalid data is input
S M O O T H I N G
0
1 5
s e c
S M O O T H I N G
1 5
0
s e c
S M O O T H I N G
I N P U T
E R R O R
Range of Input : 0 to 120sec
ex : When an invalid choice is selected
I M P - 1 3
3
< >
+ L O W
V A L U E
( x )
I M P - 1 3 3
< >
I N V A L I D
( x )
ex : When measured flow rate is over the maximum flow rate as set by + MAX VALUE parameter
(INP-132 in A-OUT) or when converted velocity e/-30m/s.
Provided that +MAX VALUE (INP-132) was 100.00,
V o l .
< A R M > B R D O
I
1 0 0 . 0 0
m
3
/ h
V o l .
< A R M > B R D O
I
F . S .
m
3
/ h
* According to each setting
Over +MAX VALUE (more than 100.01)
“F.S.” = Full Scale = Analog Output Range
ex : When values are over digits
Supposing that iMAX VALUE (INP-132) = 100 and
initial -MAX VALUE (INP-134) = -100. If +MAX VALUE is
changed to 0.300, according to the decimal places of +MAX
VALUE, a -MAX VALUE of -100.000 should be indicated.
However the maximum no. of digits of the LCD for this string is 7, including decimal point and +/-
sign. Thus -100.000, at 8 digits, will be deemed “Over Digits” and the -MAX VALUE display
indication will appear as [ ------- ].
ENT
Mark (X) indicates
invalid choice.
LCD indicates
100.00 and F.S.
alternately.
ENT
Summary of Contents for UFL-30
Page 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Page 11: ... 10 blank ...
Page 12: ...I Installation ...
Page 60: ...II Operation ...
Page 109: ...II 50 blank ...
Page 110: ...III Other ...
Page 123: ...III 14 III 2 5 Dimensions Main Unit ...
Page 128: ...III 19 blank ...
Page 129: ...III 20 blank ...
Page 145: ...III 36 blank ...
Page 165: ...III 56 blank ...