I- 39
I-2-11. Cable Connection Treatment With Chemical Binder
Chemical binder can be supplied if required
(1) Before starting cable treatment, both end of
cable sheath should be filed off elaborately by
included sandpaper to stick chemical binder
onto cable very well.
(2) Strip coaxial cable as in photo below.
(Approx. 70mm)
(3) Bundle outer shield for soldering.
(6 bundles are recommended.)
(4) Remove sheath around inner shield.
(Approx. 55mm)
(5) Bundle inner shield for soldering.
(6 bundles are recommended.)
(6) Remove Inner sheath around core cable.
(Approx. 40mm)
Note:
Pliers may be used to loosen the sheath
from the core, which makes it easier to
remove the sheath.
(7) Follow the same procedure for the other
cables.
(8) Join and solder the cable cores.
Joining only the tips of the cores is
recommended.
Joint must be double twisted as above.
(9) For certain connections, it may be suggested
to cover the cable cores using the removed