*Fasten belt
(*lsee than 100mm pipe,number of belt will be Two)
Mounting rail
Fitting plate
Transducer
Transducer holder
Screw plate
Cover
L= 300 or 450
Fig.1-2-9-24
Mounting fixture kit assembly
2-a-4. Remove transducer cover from mounting fixture kit.
The mounting fixture kit is available in two lengths, standard length of 450mm and optional length of
300mm. The 300mm-length fixture is used for small pipes of diameters less than 100mm.
2-a-5. Removal of fitting plates
The fitting plates are attached to the mounting fixture kit with 4 screws. Use a Phillips screwdriver
and unscrew and remove these screws and lift off the fitting plates (2).
2-a-6. Transducer holder
Loosen screws of the transducer holder (one side OK) to allow movement.
Fig.1-2-9-25
Fitting plate
Fitting plate
Loosen screws and
remove with fitting plate.
Loosen screws to allow
transducer movement.
Transducer
holder
Fig.1-2-9-26
Transducer holder
I- 33