1-C. Multi-Path Transducer Installation
When more than 2 pairs of transducers are employed, the basic procedures for mounting
transducers for single paths are also used in multi-path installations.
2 channel: Gauge paper should be double folded to divide the circumference by 4.
Follow the same steps used to install single path transducers.
Note: One additional pulser module is required.
4 channel: Triple fold the gauge paper to divide the circumference by 8.
Follow the same steps used to install single path transducers.
Note: Three additional pulser modules and a multi-path junction box are required.
Fig. 1-2-9-19
Fig. 1-2-9-20
P2
P1
P1
’
P2
’
I- 31