I-2-13. Input parameters by Commissioning software
Commissioning software is used to configure for the flowmeter by a personal computer.
(1) System requirements
OS : Windows 98 or later
Display: 640 x 480 or higher
Port : Serial Port EIA-232C (RS-232C) required
(NOTE: USB-serial converter available.)
(2) Software installation
Copy "FlowConfig" folder to any directory of your computer. This folder includes two software for
commissioning & application. Notice that these programs should be used by combination.
(3) Connection to main unit
CAUTION
1. Connection cable length must be less than 3 m long to comply with EC directive.
2. Do not connect the connection cable to LAN connector of the personal computer, that may
damage the flowmeter main unit and/or the personal computer.
a. Connect RJ-45 straight cable to
Digital Port 2
of the main unit.
b. Connect cable to PC by using RJ-45 / RS-232C converter.
c. Turn main unit power ON.
(4) Startup software
After connecting the PC to the flowmeter main electronics unit, double click the commissioning
software, "FlowConfig_V***.exe", to be boot up. (V*** shows the software version.)
Note: "Localize.exe" does not work independently.
RJ-45
side
1
6
5
4
3
2
8
7
1 Blue
6 Yellow
5 Green
4 Red
3 Black
2 Orange
8 White
7 Brown
D-Sub 9 Pin
side
Connector Assignment
PC
RJ-45 / RS-232C
Converter
RJ-45 Cable
(Straight Cable)
Summary of Contents for UFL-30
Page 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Page 11: ... 10 blank ...
Page 12: ...I Installation ...
Page 60: ...II Operation ...
Page 109: ...II 50 blank ...
Page 110: ...III Other ...
Page 123: ...III 14 III 2 5 Dimensions Main Unit ...
Page 128: ...III 19 blank ...
Page 129: ...III 20 blank ...
Page 145: ...III 36 blank ...
Page 165: ...III 56 blank ...