I-
8
I-2 Installation and Wiring
Installation and wiring should be worked carefully.
DANGER
Be sure to stop power supply to the main unit for installation and wiring work to prevent electrical shock .
WARNING
Make sure that wiring are made correctly as incorrect connections may result in damage to the main unit
and connected accessories.
Do not install or use the equipment in a atmosphere where any flammable or explosive gas is present.
CAUTION
After installation and wiring work, tighten the front cover screws and cable glands of the main unit exactly to
protect against water and dust.
Torque: Front cover screw
Approx. 2.5 N
・
m
Cable gland
Approx. 1.5 N
・
m
I-2-1. Setup Procedure
(1) Procedure
The basic steps in setting up the flowmeter system are outlined below.
Please note some limitations placed on operation from LCD.
No.
Step
Procedure
Reference
Chapter
From
PC
From
LCD
1
Selection of transducer position
I-2-2
------- -------
2
Installation Main Unit
Placement & Wiring connection
I-2-3 to 8
------- -------
3
Transducer Parameter Input
(1) Unit System Selection
I-2-13 (8)
Yes
N/A
(2) Q'ty of measuring path
I-2-13 (8)
(3) Pipe information input
I-2-13 (9)
(4) Transducer type
I-2-13 (9)
(5) Cable length
I-2-13 (9)
(6) Fluid parameter input
I-2-13 (9)
4
Transducer distance confirmation
I-2-13 (11)
Yes
N/A
5
Output Setup
(1) Flow rate unit settings
I-2-13 (12)
II-2-6 (2) a
Yes
Yes
(2) Totalized output settings
I-2-13 (12)
II-2-6 (4)
Yes
Yes
(3) Analog output settings
I-2-13 (13)
II-2-6 (3)
Yes
Yes
(4) Alarm setting
I-2-13 (14)
II-2-6 (3)
Yes
Yes
(5) Contact output settings
I-2-13 (15)
II-2-6 (5)
Yes
Yes
(6) Digital output settings
I-2-13 (16)
II-2-6 (6)
Yes
N/A
(7) Upload parameters
I-2-13 (17)
Yes
N/A
6
Transducer Installation
Temporary installation by using couplant.
Transducer distance parameter required.
I-2-9
------- -------
7
Cable Treatment
I-2-10 & 11
------- -------
8
Wiring Connection to Main Unit
I-2-4 to 7
------- -------
9
AGA settings
Pipe should be fulfilled with fluid.
I-2-13 (18)
Yes
Yes
10
Transducer Installation (Final)
Permanent Installation by Adhesive.
I-2-9
------- -------
11
Finalizing Setup through PC
configuration software or LCD
(1) AGA settings
(Pipe should be fulfilled with fluid.)
I-2-13 (19)
Yes
Yes
Summary of Contents for UFL-30
Page 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Page 11: ... 10 blank ...
Page 12: ...I Installation ...
Page 60: ...II Operation ...
Page 109: ...II 50 blank ...
Page 110: ...III Other ...
Page 123: ...III 14 III 2 5 Dimensions Main Unit ...
Page 128: ...III 19 blank ...
Page 129: ...III 20 blank ...
Page 145: ...III 36 blank ...
Page 165: ...III 56 blank ...