II-10
7. 2:+/-/DBL (2 system, 2 directional double range output)
Channels 1 and 2 measure forward direction and reverse direction flows respectively.
Channels 1 and 2 provide high range and low range measurements respectively.
Selectable channel switching based on flow direction or range.
Graphs of flow relative to output current are shown below.
8. SPECIAL (Special Output)
Measures forward direction flow and reverse direction flow up to 20% of forward direction flow.
Current output range is 0.8mA for -20%, 4mA for 0, and 20mA for 100%.
Same current values are output from channels 1 and 2. One range.
Graphs of flow relative to output current are shown below
正流
逆流
0
低レンジ
低レンジ
高レンジ
高レンジ
20mA
4mA
90%
80%
90%
80%
Low
Low
High
High
Backward Forward
正流
逆流
0
低レンジ
低レンジ
高レンジ
高レンジ
20mA
4mA
90%
80%
90%
80%
Low
Low
High
High
Backward Forward
正流
逆流
0
低レンジ
低レンジ
高レンジ
高レンジ
20mA
4mA
90%
80%
90%
80%
Low
Low
High
High
Backward Forward
正流
逆流
0
低レンジ
低レンジ
高レンジ
高レンジ
20mA
4mA
90%
80%
90%
80%
Low
Low
High
High
Backward Forward
正流
0
高レンジ
20mA
4mA
逆流
0.8mA
高レンジ
Hi
Hi
Backward Forward
CH1 output
Automatic double range
Direction change (FW/BW)
CH2 output
Automatic double range
Range change (HI/LO)
CH2 output
Automatic double range
CH1 output
Automatic double range
Summary of Contents for UFL-30
Page 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Page 11: ... 10 blank ...
Page 12: ...I Installation ...
Page 60: ...II Operation ...
Page 109: ...II 50 blank ...
Page 110: ...III Other ...
Page 123: ...III 14 III 2 5 Dimensions Main Unit ...
Page 128: ...III 19 blank ...
Page 129: ...III 20 blank ...
Page 145: ...III 36 blank ...
Page 165: ...III 56 blank ...