II-48
•
Example of LCD Indication
Normal
V o l .
< A R M > B R D O
I
- 1 0 . 3 4 E + 3
m
3
/ h
When the flowmeter experiences a breakdown, an error code appears in the status area of the display.
V o l .
E R R 1 6 B R D O
I
- 1 0 . 3 4 E + 3
m
3
/ h
Note:
While “Check” function is activated, no error code is indicated.
Table 4-1; Error Code List
Breakdown
Part
INPUT
PARAM.
EEP
ROM
RTC
RAM
Counter
Receiver
Trans-
mitter
Note
ID No.
32
16
8
4
2
1
B.D. = BreakDown
Error code
ERR 1
1
Transmitting
Circuit
B.D.
ERR 2
2
Receiving
Circuit
B.D.
ERR 4
4
Time Measuring Circuit B.D.
ERR 8
8
RTCRAM
B.D.
ERR 16
16
EEPROM
B.D.
ERR
32
32
Inputted Data Error
Example of Composite Error Codes
ERR 6
4
2
Time Meas. Circuit &
Receiving Circuit B.D.
ERR 9
8
1
RTCRAM &
Transmitting Circuit B.D.
ERR 20
16
4
EEPROM &
Time Meas. Circuit B.D.