1-a-7. Surface treatment
Prepare the pipe at the intended flow measurement point, making sure it is clean and free of
loose material. Sand as necessary to remove any high spots. However, be sure to maintain
the original curvature of the pipe and avoid eradicating the marks on the pipe.
1-a-8. Attachment of mounting fixture
1- Tighten the clips of the loop on the end of the
wire with the red tip.
2- Loosen the clips of the loop on the other end
of the wire (‘A’ in Fig. 1-2-9-10)
3- Adjust wire lengths so it fits the pipe circumference.
You may use gauge paper or tape measure for reference.
4- Tighten the clips on the ‘A’ side of the wires.
5- Wrap the wires around the pipe.
Clip end must be outside of the pipe, otherwise you may give damage on pipe surface.
6- Loop the ends of the wires on the turnbuckle fixture (Fig. 1-2-9-11).
Note: In case of Larger than DN1600mm, 2 turnbuckle fixtures may be used for 1 line for
easy installation.
Note: To keep parallel of each wire, you can use gauge paper as reference.
Red point
A
Wire Clip
Fig. 1-2-9-10
Fig. 1-2-9-12
Fig. 1-2-9-11
Wire-clamping
hook
Lock-Nut
Nut
Thread
Fig. 1-2-9-9
Same size of transducer surface
Polish
Installation
distance
Polish
I- 27
Summary of Contents for UFL-30
Page 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Page 11: ... 10 blank ...
Page 12: ...I Installation ...
Page 60: ...II Operation ...
Page 109: ...II 50 blank ...
Page 110: ...III Other ...
Page 123: ...III 14 III 2 5 Dimensions Main Unit ...
Page 128: ...III 19 blank ...
Page 129: ...III 20 blank ...
Page 145: ...III 36 blank ...
Page 165: ...III 56 blank ...