II-24
II-2-6 How to operate, Function by Function (LCD operation & PC operation)
This section provides a description of operation in accordance with each function that you would like
to setup. Please refer to paragraph II-1 “Functions”.
Each menu and parameter has an I.D. address in the memory as described below.
[Address: menu item]
•
The first address [ INP-1 ], this category can be reached from the menu of “MEASUREMENT” menu
.
•
The first address [ INP-2 ], this category can be reached from the menu of “INTEGRATION” menu.
•
The first address [ INP-3 ], this category can be reached from the menu of “RELAY-OUT” menu.
•
The first address [ INP-4 ], this category can be reached from the menu of “RS232C” menu.
•
The first address [ CHK ], this category can be reached from the menu of “CHECK” menu.
About using the commissioning software, please refer to the chapter I-2-13.
< Open “TOP MENU” >
In order to open “TOP MENU”, please push the
button and the
ENT
button simultaneously at any
indication. Pushing the
ESC
button allows return to the previous display (ex. [Flow Rate] Display).
< Open “INPUT” menu >
In order to open “INPUT”, please select “INPUT” and push the
ENT
button at “TOP MENU”.
Use the
&
buttons to select submenu (“MEASUREMENT”, “INTEGRATION”, “RELAY-OUT”, and
“RS232C”) and push the
ENT
button to open the selected submenu.
< Open “CHECK” menu >
In order to open “CHECK”, please select “CHECK” and push the
ENT
button at “TOP MENU”.
(1) LCD Display Relation
a. Change indicating options on the display
Default indication is [Flow Rate]. Whenever the
ENT
button is pushed, the indication will be
replaced as follows. The
ESC
button returns the indication to [Flow Rate].
* [Backward Integration] can be displayed in the case of bi-directional analog output only.
* The flowmeter resumes showing the last indication when the power was shut off.
* This display can individually indicate flow rate at every measuring path during multi-path
measuring mode.
V o l .
< A R M > B R D O
I
- 1 0 . 3 4 E + 3
m
3
/ h
P A T H 1
< A R M > B R D
I
- 1 0 . 3 4 E + 3
m
3
/ h
P A T H 2
< A R M > B R D
I
- 1 0 . 3 4 E + 3
m
3
/ h
P A T H 3
< A R M > B R D
I
- 1 0 . 3 4 E + 3
m
3
/ h
P A T H 4
< A R M > B R D
I
- 1 0 . 3 4 E + 3
m
3
/ h
ex) [Averaged Flow Rate]
is indicated by multi-path
measurement.
The
&
buttons select
individual flow rate for every
measuring path.
Push the
&
buttons to
select the number of the
measuring path.
*1 [O] = Sign of Over Range
This can be indicated during
[Averaged Flow Rate] mode.
The
ESC
button returns the
indication to [Averaged
Flow Rate] from [Individual Flow
Rate] indication.
STATUS
2
STATUS
1
Forward
Integration
Backward
Integration
Velocity
Flow
Rate
E
S
C
EN
T
EN
T
EN
T
EN
T
EN
T
EN
T
E
S
C
E
S
C
E
S
C
E
S
C
*1
Summary of Contents for UFL-30
Page 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Page 11: ... 10 blank ...
Page 12: ...I Installation ...
Page 60: ...II Operation ...
Page 109: ...II 50 blank ...
Page 110: ...III Other ...
Page 123: ...III 14 III 2 5 Dimensions Main Unit ...
Page 128: ...III 19 blank ...
Page 129: ...III 20 blank ...
Page 145: ...III 36 blank ...
Page 165: ...III 56 blank ...