II-32
e. No receiving echo (ROFF ALARM) setting
Display Menu No.
: INP-141 : ROFF ALARM
Initial value
: “0* OR” (no receiving echo from ANY path)
“1: AND” = no receiving echo from ALL paths
f. Fault tolerance setting
Display Menu No.
: INP-142 : FAULT TOLERANCE
Initial value
: “1:ON” (measurement continues as long as possible)
“0: OFF” = unconditionally halts measurement when readings are not correct (i.e. are not within
normal limits).
In case of ROFF (no receiving echo) or hardware (breakdown) problem during multi-path
measurement, output will be according to parameter, INP-143: ROFF A-OUT OPE.
In the case of a hardware (breakdown) only problem occurring during multi-path measurement,
output will be according to parameter, INP-144: B.D. A-OUT OPE.
Please refer to Chapter II-2-7 “Output operating during Aberrations in Measurement” .