III-12
Power supply AC100 to 230V +/-10% (50/60 Hz±10%)
Option : DC24V±20% (This option must be pre-selected)
Momentary
outage
AC input: 20ms, DC input: 5ms
Power
consumption
AC100V: 20VA / AC200V: 27VA
DC24V: 10W (Option)
Fuse
IEC 60127-2 SS5
Cartridge fuse-links
5.2x20 mm
Rating 2A/250V
Time-lag
High Breaking Capacity (1500A)
Rush Current Less than 20A at AC100V / Less than 32A at AC200V
Less than 30A at DC24V (Option)
Operating
temperature range
-10 to +60°C (for main unit ambient)
Storage
temperature range
-20 to +70°C
Operating
humidity range
Less than 90% RH, non-condensation
Main unit
construction
Protection Degree IP65 (IEC 60529)
Wiring
connection port
Cable gland, 7 pcs, O.D.6~12mm cable applicable
Case material
Aluminum
Coating
Melamine
Color
Munsel 10YR9. 4/0. 5
Weight
Appx. 8kg
Dimensions
260mm x 394mm x 155mm