II-14
[Totalized pulse width]
When FW INTG (forward flow totalized pulse output) or BW INTG (reverse flow totalized pulse
output) is set as relay operation output, the width of the totalized pulse (i.e. time when the relay
output is being created with the inverted output OFF) is set.
Totalized pulse width can be selected from the following.
20ms, 100ms, 500ms, 1000ms
Note
Carefully select totalized pulse width taking into consideration the totalized count up rate.
Set the totalizing unit so that the relationship between pulse width
‘a’
and pulse
interval
‘b’
is
a
b.
Pulse Width
Max. Output Rate
20ms 25
pulses/sec.
100ms 5
pulses/sec.
500ms 1
pulse/sec.
1000ms
1 pulse/2 sec.
Note
Totalized pulse is not an active output signal but a passive output signal as relay contact output..
[Setting of no receiving echo warning (ROFF ALARM)]
The relay output for no received signal warning for multi-path measurement can be set so that a
warning is generated when all paths do not receive signals (AND) or if one of the paths does not
receive a signal (OR).
[Logic inversion]
Relay logic can be inverted.
Independent ON/OFF settings can be made for each relay channel.
(
‘a’
relay operates at OFF, and
‘b’
relay operates at ON.)
a
b
a b
Summary of Contents for UFL-30
Page 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Page 11: ... 10 blank ...
Page 12: ...I Installation ...
Page 60: ...II Operation ...
Page 109: ...II 50 blank ...
Page 110: ...III Other ...
Page 123: ...III 14 III 2 5 Dimensions Main Unit ...
Page 128: ...III 19 blank ...
Page 129: ...III 20 blank ...
Page 145: ...III 36 blank ...
Page 165: ...III 56 blank ...