III-30
III-2-8 Digital output profiles (MODBUS Protocol)
(1) Overview
Flowmeter can provide digital data output of flowrate, velocity data, totalizing value as well as
measurement status.
This digital data is output from DIGITAL port1 via EIA-232 (RS-232C) connection by MODBUS
protocol when selected by commissioning software.
(2) Wiring Connection
Physical wiring between Main unit and digital output is as follows,
Main unit setting is done through Digital Port 2 by intercommunication mode under TOKYO KEIKI
original protocol.
MODBUS communication is output from Digital Port 1 in accordance with main unit setting.
CAUTION
1. Connection cable length must be less than 3 m long to comply with EC directive.
2. Do not connect the connection cable to LAN connector of the personal computer, that may
damage the flowmeter main unit and/or the personal computer.
3. Make sure that connections are made correctly as incorrect connections may result in
damage to the main unit and connected accessories.
NOTE; Each “RTS and CTS”, “DTR and DTS” of Digital Port 1 is shorted on flowmeter ‘s main unit side.
RJ-45 side
1 Blue
2.Orange
3 Black
4 Red
5 Green
6 Yellow
7 Brown
8 White
1 DCD
2.RxD
3 TxD
4 DTR
5 GND
6 DSR
7 RTS
8 CTS
D-sub 9 pin side
1 Blue
2.Orange
3 Black
4 Red
5 Green
6 Yellow
7 Brown
8 White
1
2
3
4
5
6
7
8
RJ-45 side
D-sub 9 pin side
PC with
Commissioning software
RJ-45 / RS-232C
Converter
MODBUS Output
or
D-OUT Output
RJ-45 Cable
(Straight Cable)
Summary of Contents for UFL-30
Page 1: ...Doc No KF08 001J Ultrasonic Flowmeter UFL 30 Installation Operation Manual ...
Page 11: ... 10 blank ...
Page 12: ...I Installation ...
Page 60: ...II Operation ...
Page 109: ...II 50 blank ...
Page 110: ...III Other ...
Page 123: ...III 14 III 2 5 Dimensions Main Unit ...
Page 128: ...III 19 blank ...
Page 129: ...III 20 blank ...
Page 145: ...III 36 blank ...
Page 165: ...III 56 blank ...