3.1.7 Rate (Integration Time)
Definition
Integration time is the period of time the analog-to-digital (A/D) converter takes to
sample the input signal. The integration time feature allows user to optimize either the
speed or the resolution, as well as the noise rejection and the accuracy of the
measurement. The unit of the integration time is in PLC (power line cycles). One PLC for
60 Hz is 16.67 ms, and for 50 Hz is 20 ms.
There are 4 different integration times in DL-2060 for user to select from: 0.02, 0.1, 1
and 10 PLC.
Default
For DCV, DCI, and resistance measurement, the default integration time is 1 PLC. The
user selected value is stored in a volatile memory and the default value will be restored
when the meter is power-off.
How to set the integration time:
You can set the integration time either through the front panel operation or through the
remote interface operation.
Front Panel Operation
Integration time is set indirectly when user selects the measurement resolution. Please
refer to chapter 3.1.3 for details about how to set resolution or the digits.
Remote Interface Operation
Use the following commands on your PC terminal to set the resolution.
CONFigure:< function> <range>, <resolution>
MEASure:< function>? <range>, <resolution>
SENSe:< function>:RESolution <resolution>
Or you can set integration time directly by the following commands.
SENSe:VOLTage:DC:NPLCycles {0.02|0.1|1|10|MINimum|MAXimum}