CONFigure:PERiod {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure a period measurement with the specified range and resolution. This
command does not initiate the measurement. For period measurements, the meter uses only one
“range” for all inputs between 0.33 seconds and 3.3 µsec. With no input signal applied, period
measurements return “0”.
CONFigure:CONTinuity
Preset and configure for a continuity measurement. This command does not initiate the
measurement. The range and resolution are fixed at 1kΩ and 5½ digits respectively.
CONFigure:DIODe
Preset and configure for a diode measurement. This command does not initiate the measurement.
The range and resolution are fixed at 1Vdc with 1mA current source and 5½ digits respectively.
CONFigure:TEMPerature
Preset and configure for a RTD measurement. This command does not initiate the measurement.
The resolution is fixed at 6½ digits.
CONFigure:TCOuple
Preset and configure for a thermocouple measurement. This command does not initiate the
measurement. The resolution is fixed at 6½ digits.
CONFigure?
Query a recent condition of function, and return a quoted string.