58
3.3.3 Min/Max
Definition
When the Min/Max function is enabled, the multimeter takes in a series of readings from the
measurements, stores the minimum and maximum readings in the memory, and then
calculates the average value of all readings. The number of readings taken since Min/Max
operation is enabled is recorded as well. These data are stored in a volatile memory and will
be cleared when the meter is turned off, or when the MIN/MAX is turned on, or after a
remote interface reset. The meter beeps every time when it captures a new minimum or
maximum.
How to use Min/Max
You can use the Min/Max feature either through the front panel operation or the remote
interface operation.
Front Panel Operation
User first selects the measurement function they wish to use, and then press MIN/MAX
button to enable the Min/Max function. To read the min/max value, the average value or
the readings count, user can press CONFIG button and then MIN/MAX button. Use
◁
and
▷
buttons to switch between those values. The locations of the buttons are shown
with red rectangle frames in
Figure 4-16
.
※
Note: Press MIN/MAX again to disable this feature. The “MATH” anunnciator on the
display indicates the state of a mathematical feature.
Figure 4-16
Remote Interface Operation
The following commands show you how to use the Min/Max operation from your PC
terminal.
CALCulate:FUNCtion AVERage
CALCulate:STATe {OFF|ON}