Procedure: MENU
INTERFACE
USB
ENABLE/DISABLE
Or you can do this by the other way:
Procedure: SHIFT+RATIO
OUTPUT
ENALBLE/DISABLE
Warning! You can’t use the USB interface if you want to enable the Pass/Fail signal
output. You must use the GPIB interface (G type only) or remote control. Please
disconnect the USB cable from you multimeter. The signal from the USB cable may make
the Pass/Fail signal output abnormal.
4.2 Setting Up For Remote Interface
DL-2060 may be remote controlled by USB or GPIB interface (only G type). Kind of
interface should be selected in order to use remote interface. Please refer close 3.1.9 for
detail of selection. Next, following software should be installed into the PC for control.
-Following to NI-VISA3.1
-DL Driver
Also, application software for control is prepared into the attached CD besides of above
software. Using this, an easy operation of multimeter from the PC is possible. Please
refer install manual for detail of install method.
※
Note:
Please download from our HP
or form NI
’
s HP
for latest version of each software.
4.3 Remote Interface Commands
User can instruct the multimeter to take measurements using the SCPI commands after
the appropriate setup for their selected remote interface.
The following conventions are used in SCPI command syntax.
Triangle brackets (<>)
indicates that you must specify a value for the enclosed
parameter.
The square brackets ([])
indicates that the parameter is optional and can be