54
The internal settings of automatic trigger delay
The delay time for automatic trigger delay is determined by the selection of
measurement function, range, and the setting of integration time and AC filter speed.
The following table shows the default delay time for each measurement
setting.
Table 4-4
Measurement Function
Setting
Trigger Delay Time
DCV / DCI
PLC >= 1
1.5 ms
PLC < 1
1.0 ms
Ω2 and Ω4
(PLC >= 1)
100Ω to 100kΩ
1.5 ms
1 MΩ
15 ms
10 MΩ to 100 MΩ
100 ms
Ω2 and Ω4
(PLC < 1)
100Ω to 100kΩ
1.0 ms
1 MΩ
10 ms
10 MΩ to 100 MΩ
100 ms
ACV / ACI
(Remote Interface/External
Trigger/Single Trigger)
3 Hz
7.0 s
20 Hz
1.0 s
200 Hz
600 ms
ACV / ACI
(Front Panel w/ Auto Trigger
On)
3 Hz
1.5 s
20 Hz
200 ms
200 Hz
100 ms
Frequency/Period
Remote Interface /
ternal
1.0 s
Front Panel
/
Auto Trigger ON
0 s
How to specify a delay time
User can set the delay time from either the front panel operations or the remote
interface operations.
Front Panel Operation
To set the delay time, press MENU button and use
◁
and
▷
buttons to locate “TRIG”
option. Press ENTER to select it. Then again find “DELAY” option using
◁
and
▷
buttons. Press ENTER to select it. Then use
◁
and
▷
buttons to move through the
digits and
△
and
▽
buttons to set the desired delay time (between 0 to 3600 second).
The locations of the buttons are shown with red rectangle frames in
Figure 4-13
.