INSTRUCTION MANUAL

KEW MATE  2000/2001

DIGITAL MULTIMETER 

WITH 

AC/DC CLAMP SENSOR

KYORITSU ELECTRICAL INSTRUMENTS

WORKS, LTD.

1.SAFETY WARNINGS

This instrument has been designed and tested according to IEC Publication 61010: 

Safety Requirements for Electronic Measuring Apparatus.  This instruction manual 

contains warnings and safety rules which must be observed by the user to ensure 

safe operation of the instrument and to retain it in safe condition.  Therefore, read 

through these operating instructions before starting using the instrument.

 

   

WARNING

● Read  through  and  understand  instructions  contained  in  this  manual  before 

starting using the instrument.  

● Save  and  keep  the  manual  handy  to  enable  quick  reference  whenever 

necessary. 

● Be  sure  to  use  the  instrument  only  in  its  intended  applications  and  to  follow 

measurement procedures described in the manual. 

● Be sure to understand and follow all safety instructions contained in the manual.

Failure  to  follow  the  above  instructions  may  cause  injury,  damage  to  the  instrument 

and/or damage to equipment under test.

The symbol   indicated on the instrument means that the user must refer to related 

parts of the manual for safe operation of the instrument.  Be sure to carefully read the 

instructions following each   symbol in this manual.

  

DANGER

   is reserved for conditions and actions that are likely to cause serious or fatal 

injury.

  

WARNING

 is reserved for conditions and actions that can cause serious or fatal injury.

  

CAUTION

   is  reserved  for  conditions  and  actions  that  can  cause  minor  injury  or 

instrument damage.

Following symbols are used on the instrument and in the instruction manual. Attention 

should be paid to each symbol to ensure your safety.

Refer to the instructions in the manual.

This symbol is marked where the user must refer to the instruction manual so as not 

to cause personal injury or instrument damage.
Indicates an instrument with double or reinforced insulation.
Indicates that this instrument can clamp on bare conductors when measuring

a voltage corresponding to the applicable Measurement category, which is marked 

next to this symbol.
Indicates AC (Alternating Current).
Indicates DC (Direct Current).
Indicates AC and DC.

   

DANGER

● Never  make  measurement  on  circuits  with  a  maximum  voltage  difference  of 

600VAC/DC or greater between conductors (300VAC/DC or greater between a 

conductor and ground). 

● Do  not  attempt  to  make  measurement  in  the  presence  of  flammable  gasses.  

Otherwise,  the  use  of  the  instrument  may  cause  sparking,  which  leads  to  an 

explosion.  

● Never attempt to use the instrument if its surface or your hand is wet.  

●Do not exceed the maximum allowable input of measuring ranges.

● Never open the battery compartment cover while making measurement.

● Never  try  to  make  measurement  if  any  abnormal  conditions,  such  as  broken 

Transformer jaws or case is noted.

● The  instrument  is  to  be  used  only  in  its  intended  applications  or  conditions. 

Otherwise,  safety  functions  equipped  with  the  instrument  doesn t  work,  and 

instrument damage or serious personal injury may be caused.

   

WARNING

● Never attempt to make any measurement, if any abnormal conditions are noted, 

such as broken case, cracked test leads and exposed metal parts.   

● Do not turn the Function Selector Switch while the test leads are connected to 

the circuit under test. 

● Do not install substitute parts or make any modification to the instrument.  Return 

the instrument to Kyoritsu or your distributor for repair or re-calibration.

● Do not try to replace the batteries if the surface of the instrument is wet. 

● Always disconnect the clamp sensor and the test leads from the circuit under test 

and  switch  off  the  instrument  before  opening  the  battery  compartment  cover  for 

battery replacement.

   

CAUTION

●  Make  sure  that  the  Function  Selector  Switch  is  set  to  an  appropriate  position 

before making measurement.

●  Always make sure to place the test leads in the test lead holder before making 

current measurement. 

●  Do not expose the instrument to the direct sun, extreme temperatures or dew fall.  

●  Be  sure  to  set  the  Function  Selector  Switch  to  the  "OFF"  position  after  use. 

When the instrument will not be used for a long period of time, place it in storage 

after removing the batteries.

●  Use  a  damp  cloth  and  detergent  for  cleaning  the  instrument.    Do  not  use 

abrasives or solvents.

Measurement Category:

    To  ensure  safe  operation  of  measuring  instruments,  IEC  61010  establishes  safety 

standards for various electrical environments, categorized as O to CAT.IV, and called 

measurement  categories.  Higher-numbered  categories  correspond  to  electrical 

environments  with  greater  momentary  energy,  so  a  measuring  instrument  designed 

for  CAT.III  environments  can  endure  greater  momentary  energy  than  one  designed 

for CAT.II.

   O 

:  Circuits which are not directly connected to the mains power supply. 

   CAT.II  :  Electrical circuits of equipment connected to an AC electrical outlet by a 

power cord.

   CAT.III  :  Primary  electrical  circuits  of  the  equipment  connected  directly  to  the 

distribution panel, and feeders from the distribution panel to outlets.

   CAT.IV  :  The circuit from the service drop to the service entrance, and to the power 

meter and primary over-current protection device (distribution panel).

2.FEATURES

● Permits  AC/DC  current  measurement  up  to  60A  using  a  clamp  sensor  that  comes  

standard with the instrument

●Clamp sensor for ease of use in crowded cable areas and other tight places

● Permits current measurement with an open current-clamp sensor that does not require 

opening and closing operations by the user

●Auto-power-save function

●Buzzer for easy continuity checking

●Data hold function to freeze the readings

●LCD with a 3400 count full scale bar graph

●Shock absorbing holster for ease of storage

● Designed  to  international  safety  standard  IEC61010-1:  measurement  category 

CAT.

, 300V and pollution degree 2.

O: Device which is 

     not directly 

     connected to the 

     mains power supply

3.SPECIFICATIONS

●Measuring Ranges and Accuracy (at 23℃±5℃, relative humidity75% or less)

※Electromagnetic RF field  ≦ 1 V/m
  compatibility     ACV/DCV/OHMS/FREQUENCY total accuracy = specified accuracy
  (IEC 61000-4-3)     ACA/DCA            total accuracy = specified a5dgt
          RF  transmitters  such  as  mobile  telephones  may  not  be  used  in  close 

proximity.

●Safety Standard 

IEC 61010-1

   

 

measurement CAT.

, 300V, pollution degree 2

   

 

measurement CAT.

, 600V, pollution degree 2

   

 

IEC 61010-031

   

 

IEC 61010-2-032, IEC 61010-2-033

   

 

IEC 61326-1 (EMC), EN 50581 (RoHS)

●Operating System 

Dual integration

●Display 

 

 Liquid  crystal  display  with  maximum  reading  of  3399  as 

well as units and annunciators

   

 

Bar graph with maximum points of 33

●Over Input Indication 

"OL" on the LCD (Ω ranges only)

●Auto-ranging Operation 

 Shifts to the next higher range when bar graph increases 

to 33 points

   

 

 Shifts to the next lower range when bar graph decreases 

to 3 points

●Sample Rate  Numeric reading: about 400ms, 

   

 

bar graph: about 20ms

MODEL

2000

2001

Range

 60A

100A

AC Current   A                                                 

DC Current   A                                                  

AC Voltage  V Input impedance: 10MΩ

DC Voltage  V  Input impedance: 10MΩ

Measuring Range

0-60.0A

0-100.0A

Accuracy

 ± 2.0%rdg ± 5dgt(50/60Hz)

 ± 2.0%rdg ± 5dgt(50/60Hz)

MODEL

2000

2001

Range

 60A

100A

Measuring Range

0- ± 60.0A

0- ± 100.0A

Accuracy

± 2.0% rdg ± 5dgt

± 2.0% rdg ± 5dgt

Range

 3.4V

 34V

 340V

 600V

Measuring Range

 0-600V

(Auto-ranging)

Accuracy

± 1.5%rdg ± 5dgt(50-400Hz)

Range

 340mV

 3.4V

 34V

 340V

 600V

Measuring Range

0- ± 600V

(Auto-ranging)

Accuracy

± 1.5%rdg ± 4dgt

Resistance Ω/ 

Range

 340 Ω

 3.4k Ω

 34k Ω

 340k Ω

 3.4M Ω

 34M Ω

Measuring Range

0-33.99M Ω

 (Auto-ranging)

Accuracy

± 1.0%rdg ± 3dgt

Buzzer beeps below 30 ± 10Ω

(Continuity  buzzer  works  on  340Ω  range 

only)

± 5%rdg ± 5dgt

± 15%rdg ± 5dgt

Frequency Hz 

Range

Current

Voltage

Measuring Range

0-3.399kHz

3.4kHz-10kHz

 (Auto-ranging)

0-3.399kHz

3.4kHz-33.99kHz

34kHz-300kHz

 (Auto-ranging)

Accuracy

± 0.1%rdg ± 1dgt

± 0.1%rdg ± 1dgt

●Location for use 

 Indoor use, Altitude up to 2000m

●Accuracy-insured 

23℃±5℃, relative humidity 75% or less 

 Temperature and 

(without condensation)

 Humidity Ranges

●Operating Temperature 

0-40℃, relative humidity 85% or less 

 and Humidity Range  

(without condensation)   

●Storage Temperature 

-20-60℃, relative humidity 85% or less

 and Humidity Range  

 (without condensation)

●Source 

 

Two 1.5VDC R03 (UM-4) batteries

●Current Consumption 

Approx. 10mA

●Power-save Function 

 Shifts to the power-save state about 10 minutes after the 

last switch operation

   

 

 (current consumption: approx. 10μA)   

●Overload Protection 

AC/DC current ranges:  MODEL2000  AC/DC  72A  for  10 

seconds

            

 AC/DC current ranges:  MODEL2001  AC/DC 

120

A  for 

10 seconds

            

 AC/DC voltage ranges: AC/DC 720V for 10 seconds

   

 

Resistance ranges: AC/DC 720V for 10 seconds

   

 

Frequency ranges: AC/DC 720V for 10 seconds

●Withstand Voltage 

 AC3470V  for  5  sec.  between  electrical  circuit  and 

housing case

●Insulation Resistance 

 10MΩ or greater at 1000V between electrical circuit and 

housing case

●Conductor Size 

MODEL2000 Approx.  6mm diameter max.

            

 MODEL2001 Approx. 10mm diameter max. 

●Dimensions  MODEL2000 128(L)×87(W)×24(D)mm

            

 MODEL2001 128(L)×92(W)×27(D)mm 

●Weight 

 

MODEL2000 Approx. 210g

            

 MODEL2001 Approx. 220g

●Accessories Two R03 (UM-4) batteries

   

 

Instruction Manual

4.INSTRUMENT LAYOUT

Data Hold Button

MODEL2001 MODEL2000

Range Hold Button

Test Lead

LCD Display

AC/DC Clamp Sensor

DCA

Zero Adjust

Knob

Digital Multimeter

Function Selector Switch

Holster

Barrier

Barrier

AC/DC

Battery Voltage Warning

Auto Range

Data Hold

Unit

Bar Graph

Buzzer

Manual

Range

Wind the lead around the folder.

Test Leads

AC/DC Clamp Sensor

5.PREPARATIONS FOR MEASUREMENT

(1)Checking battery voltage

Set the Function Selector Switch to any position other than the OFF position. 

 If the marks on the display is clearly legible without symbol 

"BATT"

 showing, battery 

voltage  is  OK.  If  the  display  blanks  or 

"BATT"

  is  indicated,  replace  the  batteries 

according to section 

8: Battery Replacement

.

NOTE

When the instrument is left powered on, the auto-power-save function automatically 

shut the power off; The display blanks even if the Function Selector Switch is set to 

a  position  other  than  the  OFF  position  in  this  state.    To  power  on  the  instrument, 

turn the Function Selector Switch or press the Data Hold Button.  If the display still 

blanks, the batteries are exhausted.  Replace the batteries.

(2)Make sure that the Function Selector Switch is set to the appropriate range.

 

Also  make  sure  that  data  hold  function  is  not  enabled.    If  inappropriate  range  is 

selected, desired measurement cannot be made.

(3)Install Test Lead to the Holster on the side of body

 

It is possible to measure with seeing the LCD Display keep 

Test Lead installing to the Holster.

6.HOW TO MAKE MEASUREMENT

6-1 Current 

Measurement

   

DANGER

● In order to avoid possible shock hazard, never make measurement on circuits with 

a  maximum  voltage  difference  of  600VAC/DC  or  greater  between  conductors 

(300VAC/DC or greater between a conductor and ground). 

● Do not make measurement with the test leads connected to the circuit under test. 

Never make measurement with the battery compartment cover removed.

●Keep your fingers and hands behind the barrier during measurement.

 

   

CAUTION

● When handling the clamp sensor, exercise caution not to apply excessive shocks 

or vibration to the sensor. 

● 

M

aximum  measurable  conductor  size  is  MODEL2000  6mm  /  MODEL2001 

10mm in diameter.

AC/DC

Battery Voltage Warning

Auto Range

Data Hold

Unit

Bar Graph

Buzzer

Manual

Range

Wind the lead around the folder.

Test Leads

AC/DC Clamp Sensor

 6-1-1 DC Current Measurement

(1)The Function Selector Swith to the "  A" position.

 ("DC" and "AUTO" marks are shown on the top of the display.)

(2) Turn the 0(Zero) ADJ knob to set the reading of the multimeter to zero. (If this zero 

adjustment is made incorrectly, measurement errors will result.) 

(3)  Adjust one of the conductors to the center of the clamp sensor's arrow. 

  (  When  the  position  of  the  conductor  is  not  at  the  center  of  the  arrow,  the  error 

occurs.)

     Measured value is shown on the display.

Note:  When current flows from the upside to the underside of the instrument, the polarity 

of the reading is positive (+).  Otherwise, the polarity of the reading is negative (-). 

 6-1-2 AC Current Measurement

(1)Set the Function Selector Switch to "  A."

 ("AC" and "AUTO" marks are shown on the top of the LCD.)

(2) Adjust one of the conductors to the center of the clamp sensor's arrow.

   (  When  the  position  of  the  conductor  is  not  at  the  center  of  the  arrow,  the  error 

occurs.)

     Measured value is shown on the display.

Note:  Unlike DC current measurement, zero adjustment is not necessary.  There is not 

polarity indication either.

6-2 Voltage Measurement

   

DANGER

● In  order  to  avoid  possible  shock  hazard,  never  make  measurement  on  circuits 

with  a  maximum  voltage  difference  of  600VAC/DC  or  greater  between 

conductors (300VAC/DC or greater between a conductor and ground). 

● Do not make measurement with the battery compartment cover removed.

●Keep your fingers and hands behind the barrier during measurement.

 6-2-1 DC Voltage Measurement

(1)Set the Function Selector Switch to "  V."

  ("DC" and "AUTO" marks are shown on the top of the LCD.)

(2) Connect the red test lead to the positive (+) side of the circuit under test and the 

black test lead to the negative (-) side.  Measured voltage value is shown on the 

display.

  When the connection is reversed, "-" is shown on the display.

 6-2-2 AC Voltage Measurement

(1)Set the Function Selector Switch to "  V."

  ("AC" and "AUTO" marks are shown on the LCD.)

(2) Connect the test leads to the circuit under test.

  Measured voltage value is shown on the display.

6-3 Resistance Measurement                                                

   

DANGER

● Never make measurement on circuits that are live. 

● Never make measurement with the battery compartment cover removed. 

●Keep your fingers and hands behind the barrier during measurement.

Conductor

MODEL2000

MODEL2001

Adjust a

conductor to

the center of

the arrow.

Correct

Wrong

(1)Set the Function Selector Switch to "Ω/  ."

(2) Check that the display shows over-range. Short the test leads and check that the 

buzzer beeps and the display reads zero.

(3) Connect  the  test  leads  to  the  circuit  under  test.  Measured  resistance  value  is 

shown on the display.  When the measured value is below about 30Ω, the buzzer 

beeps.

Note:  When  the  test  leads  are  shorted,  the  display  may  read  a  small  resistance  value.  

This is the resistance of the test leads.  

             If there is an open in either of the test leads, "OL" is shown on the display. 

       On the 340Ω range, "  " is shown on the left side of the LCD.

6-4 Frequency Measurement

   

DANGER

●  In order to avoid possible shock hazard, never make measurement on circuits with 

a  maximum  voltage  difference  of  600VAC/DC  or  greater  between  conductors 

(300VAC/DC or greater between a conductor and ground). 

●  Do not make measurement with the test leads connected to the circuit under test. 

Never make measurement with the battery compartment cover removed.

●  Do  not  make  current  measurement  with  the  test  leads  connected  to  the  circuit 

under test.

●Keep your fingers and hands behind the barrier during measurement.

(1)Set the Function Selector Switch to "Hz."

(2) Measuring frequency of current:

   Adjust one of the conductors to the center of the clamp sensor's arrow. Measured 

value is shown on the display.

  Measuring frequency of voltage:

   Connect the test leads to the circuit under test.  Measured frequency is shown on 

the display.

Note:  Measuring range of current frequency is 0-10kHz with minimum measurable input 

of MODEL2000 15A(Typ)/MODEL2001 25A(Typ). 

         Measuring  range  of  voltage  frequency  is  0-300kHz  with  minimum  measurable 

input of 30V(Typ).

     When measuring frequency, do not attach the clamp sensor and the test leads to 

the circuit under test simultaneously.

7.OTHER FUNCTIONS

7-1  Auto-Power-Save Function                                         

NOTE

A small amount of current is consumed even in the power-save state.  Make sure to 

set  the  Function  Selector  Switch  to  the  OFF  position  when  the  instrument  is  not 

used.

Prohibition

This  function  helps  to  avoid  unwanted  exhaustion  of  the  batteries  because  of 

leaving  the  instrument  powered  on  and  extend  battery  life.    The  instrument 

automatically  shifts  to  the  power-save  state  about  10  minutes  after  the  last 

Function Selector Switch or other switch operation.  

To  return  to  the  normal  state:    Turn  the  Function  Selector  Switch  or  press  the 

Data  Hold  Button  twice  to  exit  the  power-save  state  and  enable  measurement 

functions.

7-2 Data Hold Function

This  is  a  function  to  freeze  a  measured  value  on  the  display.    Press  the  Data 

Hold Button once to hold the current reading.  In this data hold state, the reading 

is held even if input varies.  "H" and "  " marks are shown on the LCD instead 

of "AUTO" mark.  

To exit the data hold state, press the Data Hold Button again.

7-3 Range Hold Function

The  instrument  defaults  to  auto-ranging  ("AUTO"  is  shown  on  the  LCD).  

Pressing  the  Range  Hold  Button  enables  manual  selection  among  measurement 

ranges("  " mark is shown on the LCD instead of "AUTO" mark)

Press the Range Hold Button to select a higher range.  

To  switch  from  manual  range  selection  to  auto-ranging,  press  down  the  Range 

Hold  Button  for  about  one  seconds,  or  turn  the  Function  Selector  Switch  to 

another position before setting it back to the current range. 

8.BATTERY REPLACEMENT

   

WARNING

● In order to avoid possible shock hazard, always disconnect the test leads from 

the circuit under test and set the Function Selector Switch to the OFF position 

before trying to replace the batteries.

  

 CAUTION

● Do not mix new and old batteries. 

● Install  batteries  in  the  orientation  as  shown  inside  the  battery  compartment, 

observing correct polarity.

When the battery voltage warning mark "BATT" is shown on the top left corner of the 

LCD,  replace  the  batteries.    Note  that  the  display  blanks  and  "BATT"  mark  is  not 

shown if the batteries are completely exhausted.

 (1) Set the Function Selector Switch to "OFF."

 (2) Remove the instrument from the holster.

 (3) Loosen  the  battery-compartment-cover-fixing  screw  on  the  lower  back  of  the 

instrument. 

 (4) Replace the batteries with two new R03 (UM-4) 1.5V batteries.

 (5) Put the battery compartment cover back in place and tighten the screw.

Screw

Battery

Compartment

Cover

Batteries

92-1507E

8-16

DISTRIBUTOR

Kyoritsu  reserves  the  rights  to  change  specifications  or 
designs  described  in  this  manual  without  notice  and  without 
obligations.