38
B.
Press CONFIG and then select a function as you wish from DCV, DCI (SHIFT +DCV),
Ω2, Ω4 (SHIFT + Ω2), FREQ and PERIOD (SHIFT + FREQ).
Press ENTER on RESOLUTION submenu. If it is not shown on the display, use
◁
and
▷
to locate it.
Use
◁
and
▷
to find your desired resolution. Press ENTER on your choice.
※
Note: The resolution of AC measurements is fixed at 6 1/2 digit.
※
Note: The options for frequency and period measurements are 4 1/2 (slow), 5 1/2
(slow) and 6 1/2 (slow).
Figure 4-3
Remote Interface Operation
Use the following commands on your PC terminal to set the resolution for your
measurement.
CONFigure:<function> <range>,<resolution>
MEASure:<function>? <range>,<resolution>
SENSe:<function>:RESolution <resolution>
Users can specify the resolution in the same units as the measurement function, not in
number of digits. For instance, specify the resolution in volts and the frequency in herts.
The examples are in the following.
CONF:VOLT:DC 10,0.001 (4 1/2 digits on the10Vdc range)
MEAS:CURR:AC? 1,1E-6 (6 1/2 digits on the1A range)
CONF:FREQ 1kHz,0.1Hz (1000 Hz input, 0.1Hz resolution)
VOLT:AC:RES 0.05 (50mV resolution on the ac function)
or