2.5 Continuity Measurements
DL-2060 uses 1 K Ω range for the continuity measurement. The meter beeps when the
test resistance is less than the threshold resistance. The default threshold resistance is
10Ω, but user can set the threshold resistance to anything between 1 Ω and 1 K Ω. The
resistance value set by user is stored in a volatile memory and will be cleared after the
meter has been turned off. The source current for the continuity measurement is 1 mA.
Warning! The maximum input voltage allowed is 1000 V
(peak). Applying excess
voltage may damage the meter and cause unpredictable hazards.
How to measure the continuity
Selects input signal connections on front or rear panel.
①
Connect the test leads to the terminals as shown in
Figure 3-2
.
②
Set REConnect the input signal as the following figure.
③
Set threshold resistance via pressing CONT buttons. (Or skip this step if
default resistance is to be used) When ready, press ENTER button.
④
Press CONT button.
⑤
The measured value will be shown on the display automatically. Meter “beeps” when
measured resistance value is lower than threshold value.
Figure 3-10
1
3
4
5
2