65
Figure 4-22
shows the locations of these buttons with red rectangle frames
.
Figure 4-22
Remote Interface Operation
Use the following commands from your PC terminal to make the dB measurement:
CALCulate:FUNCtion DB
CALCulate:STATe {OFF|ON}
CALCulate:STATe?
CALCulate:DB:REFerence {<
value
>|MINimum|MAXimum}
B. dBm
Definition
With dBm selected, a voltage measurement is displayed as the level of power, relative to
1 milliwatt, dissipated through a reference resistance. The reference resistance is
adjustable in DL-2060. The calculation of dBm is defined as below:
mW
Z
V
dB
ref
in
m
1
/
log
10
2
※
Note:
in
V
is the input voltage.
ref
Z
is the reference resistance.
※
Note: Change of reference resistance will not affect the stored relative reference value.
The Zref is adjustable with a range from 50Ω to 8000 Ω. The default value is 600 Ω.
The user selected reference value is stored in a volatile memory and will be cleared
after the multimeter has been power-off.
This feature is available for DCV and ACV only.
How to set the reference resistance
User can set the reference resistance either through the front panel operation or the
remote
interface operation.
Front Panel Operation
Press CONFIG + SHIFT + dBm(MIN/MAX), then use
◁
and
▷
to move through the
digits and
△
and
▽
to increase or decrease the numbers to a desired value. Press
ENTER to set the value. The locations of the buttons are shown with red rectangle frames
in
Figure 4-23
.
or
Summary of Contents for DL-2060