Error Message
Errors are retrieved in first-in-first-out (FIFO) order. The first error returned is the first
error that was stored. When user has read all errors from the queue, the
ERROR
annunciator turns off. DL-2060 beeps once each time an error occurs.
Should more than 20 errors have existed, the last error stored in the queue (the most
recent error) is replaced with -350, “Too many errors”. No additional errors are saved
until you remove errors from the queue, the multimeter responds with +0, “No error”.
The error queue is cleared when power has been off or after a *CLS (clear status)
command has been executed. The *RST (reset) command does not clear the error
queue.
5.1 Error Type
There is only one type over here, “execution errors”.
Please refer to 3.4.12 for errors arised by selftest.
5.1.1 Execution Errors
-101 Invalid character
An invalid character was found in the command string.
-102 Syntax error
Invalid syntax was found in the command string.
-103 Invalid separator
An invalid separator was found in the command string.
-104 Data type error
A parameter type error was found in the command string.
-105 GET not allowed
A Group Execute Trigger (GET) is not allowed in the command string.