73
please note that the channel 1 is relative to the channel 6, channel 2 to channel 7 (or
channel 1 is relative to the channel 11, channel 2 to channel 12) and so on.
※
Note: Setting range of scanner card is only limited on DCV, DCI, ACV, ACI, Frequency, Period,
Resistance ,and RTD
functions. OP-41T also supports Thermocouple
measurement.
As to the
current measurements, users can install current shunts on the circuit board to allow indirect
AC and DC current measurements to be made through current shunts only.
Remark 1:
In the default situation, current shunts are not installed in the scanner card.
Remark 2:
When the current shunts are installed the channels will be limited to the current
measurement unless users release the shunts.
※
Note: The Information regarding the Limitation of Input Signals in Scanner Card is in the
following statement.
10 Channels: 10 channels of 2-pole relay input. All channels configurable to 4-pole.
CAPABILITIES: Multiplex one of ten 2-pole or one of five 4-pole signals into DMM.
20 Channels: 20 channels of 2-pole relay input. All channels configurable to 4-pole.
CAPABILITIES: Multiplex one of twenty 2-pole or one of ten 4-pole signals into DMM.
MODEL
OP-41
OP-41T
OP-42
Maximum AC Voltage
125V rms or 175V peak, 100kHz, 1A
switched, 62.5VA (resistive load)
110V rms or 155V peak, 100kHz, 1A
switched, 30VA (resistive load)
Maximum DC Voltage
110V, 1A switched, 30VA (resistive load)
Number of Channels
10
10
20
Contact Life
>100000 operations at maximum signal level; >100000000 operations cold switching.
Contact Resistance
<1ohm at end of contact life
Actuation Time
5ms maximum on/off
Contact Potential
<±500nV typical per contact, 1μV max
<±500nV typical per contact pair, 1μV max
Connector Type
Screw terminal, from #22 to 28 AWG wire size
Isolation btw Any Two terminals
>10 Gohm, < 75pF
Isolation btw Any Terminal and Earth >10 Gohm, < 150pF
Common Mode Voltage
350V peak btw any terminal and earth
200V peak btw any terminal and earth
Max. Voltage btw Any Two Terminals 200V peak
160V peak
Max. Voltage btw Any Terminal and
DL-2060 Input LO
200V peak
160V peak
DCI and ACI channels
CH1,CH6
All channels
CH1, CH11
Shunt installation
R1 and R2 on PCB
All channels
J21, J23
※
Please use thin covering wire if it is tight in the cover case.