110
MEASure:CURRent:AC? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a AC current measurement with the specified range and resolution. The reading
is sent to the output buffer. For AC measurement, resolution is fixed at 6 ½ digits. Therefore the
resolution parameter only affects the front panel display.
MEASure:RESistance? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a 2-wire Ω measurement with the specified range and resolution. The reading is
sent to the output buffer.
MEASure:FRESistance? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a 4-wire Ω measurement with the specified range and resolution. The reading is
sent to the output buffer.
MEASure:FREQuency? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a frequency measurement with the specified range and resolution. The reading
is sent to the output buffer. For frequency measurements, the meter uses only one “range” for all
inputs between 3Hz and 300kHz. With no input signal applied, frequency measurements return
“0”.
MEASure:PERiod? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a period measurement with the specified range and resolution. The reading is
sent to the output buffer. For period measurements, the meter uses only one “range” for all
inputs between 0.33 seconds and 3.3
μ
sec. With no input signal applied, period measurements
return “0”.
MEASure:CONTinuity?
Preset and make a continuity measurement. The reading is sent to the output buffer. The range
and resolution are fixed at 1kΩ and 5 ½ digits respectively.
MEASure:DIODe?
Preset and make a diode measurement. The reading is sent to the output buffer. The range and
resolution are fixed at 1Vdc with 1mA current source and 5 ½ digits respectively.
MEASure:TEMPerature?
Preset and make a temperature measurement with RTD. The reading is sent to the output buffer.
MEASure:TCOuple?
Preset and make a temperature measurement with thermocouple. The reading is sent to the
output buffer.
Summary of Contents for DL-2060
Page 37: ...31 Figure 3 17 1 2 3 5 5 4...
Page 136: ...130...
Page 137: ...131...
Page 138: ...132 D DL 2060 s Dimension...