145
G. Notice of Scanner Card
Scanner Command
OP-41
OP-42
OP-41T
ROUTe:CLOSe <channel>
〇
〇
ROUTe:CLOSe?
〇
〇
ROUTe:OPEN
〇
〇
ROUTe:STATe?
〇
〇
ROUTe:SCAN:FUNCtion<channel>,{<function>|“VOLT:DC”|“VOLT:AC”|“FREQuency”|
“RESistance”|“FRESistance”|“PERiod”|
“
TEMPerature
”| “NONE”}
※
〇
ROUTe:SCAN:FUNCtion<channel>,{<function>|“VOLT:DC”|“VOLT:AC”|“FREQuency”|
“RESistance”|“FRESistance”|“PERiod”|“
TCOuple
”| “
TEMPerature
”| “NONE”}
※
〇
ROUTe:SCAN:FUNC? <channel>
〇
〇
ROUTe:SCAN:TIMER?
〇
〇
ROUTe:SCAN:TIMER <value>
〇
〇
ROUTe:SCAN:COUNT?
〇
〇
ROUTe:SCAN:COUNT <value>
〇
〇
ROUTe:SCAN:STATe?
〇
〇
ROUTe:SCAN:SCAN
〇
〇
ROUTe:SCAN:STEP
〇
〇
[SENSe:]TCOuple:RJUNction:RSELect {REAL|SIMulated }
※
〇
[SENSe:]TCOuple:RJUNction:RSELect?
※
〇
[SENSe:]TCOuple:RJUNction:REAL?
※
〇
[SENSe:]TEMPerature:RTD:TYPE
{PT100|D100|F100|PT385|PT3916|USER|SPRTD|NTCT}, @{scanner channel
number}
※
〇
〇
[SENSe:]TEMPerature:RTD:TYPE? @{scanner channel number}
※
〇
〇
[SENSe:]TEMPerature:TRANsducer FRTD, @{scanner channel number}
※
〇
〇
[SENSe:]TEMPerature:TRANsducer RTD, @{scanner channel number}
※
〇
〇
[SENSe:]TEMPerature:TRANsducer? @{scanner channel number}
※
〇
〇
[SENSe:]TCOuple:TYPE {E|J|K|N|R|S|T} , @{scanner channel number}
※
〇
[SENSe:]TCOuple:TYPE? @{scanner channel number}
※
〇
〇
:
Available
、
※
:
The bold figure
means that it is available with firmware ver. 1.05 or upward.
Operation with firmware version 1.05 or upward
CONFIG
TCOUPL
TYPE
UNITS
RJUNCTION
REAL
SIMULATED
VIEW REAL
E, J, K, N, R, S, T
℃, °
F, K
While installing the TC-Scanner Card(OP-41T) in DL-2060
The TCOUPL function on the front panel will be disenabled automatically. You can only use
the Card for thermocouple measurement.
You can only use REAL mode for RJUNCTION.
While installing the Scanner Card(except OP-41T) in DL-2060
REAL and VIEW REAL functions are unavailable.