50
C. Single Trigger Mode (accessible only through the front panel)
Definition
Single trigger mode takes one reading (or specified number of readings) each time when
user presses SINGLE key. (Please refer to 3.2.3 for setting the number of samples on
each trigger.) When the TRIG annunciator on the display is lit, the meter is ready for
next trigger event. The single trigger mode is only available through the front panel
operations.
How to use Single Trigger
First select the measurement function, and then press SINGLE key on the front panel to
set the trigger mode. The location of the button is shown with a red rectangle frame in
Figure 4-10
.
Figure 4-10
3.2.2 Trigger Source
In DL-2060, user can specify the trigger source to be one of these three options: front
panel operations, external hardware trigger source and remote interface operations.
Front Panel Operation
Use the front panel buttons - AUTO TRIGGER for auto triggering and SINGLE for single
triggering.
External Hardware
User applies a trigger pulse to the EXT TRIG (BNC) terminal on the rear panel. External
hardware triggering is like using single trigger but the trigger source is an external
hardware. When the multimeter receives a pulse from the trigger source, it takes one
reading, or a specified number of readings. To set the external hardware trigger, press
both the SHIFT + SINGLE buttons to enable external hardware trigger. The lit “EXT”