-170~-178 Expression errors
The meter does not accept mathematical expression.
-211 Trigger ignored
A Group Execute Trigger (GET) or *TRG was received but the trigger was ignored.
-213 Trigger deadlock
A trigger deadlock occurs when the trigger source is BUS and a READ? Command is
received
-214 Init Ignored
An INITiate command was received but could not be executed because a measurement
was already in progress. Send a device clear to halt a measurement in progress and
place the meter in the “idle” state.
-221 Settings conflict
This error can be generated in one of the following situations:
Situation 1:
You sent a CONFigure or MEASure command with autorange enabled and
with a fixed resolution.
Situation 2:
You turned math on and then changed to a math operation that was not
valid with the present measurement function.
-222 Data out of range
A numeric parameter value is out of range.
-223 Too much data
A character string was too long.
-224 Illegal parameter value
A discrete parameter was received which was not a valid choice for the command.
-230 Data Stale
A FETCh? Command was received but the memory was empty.
-350 Too many errors
The error queue is full.
-410 Query INTERRUPTED
A command was received which sends data to the output buffer, but the output buffer
contained data from a previous command.