3.4.9 Initial Mode
The section contains two selections: “DEFAULT” and “SAVE DATA”. You can select “SAVE
DATE” to save the current configuration or select “DEFAULT” to restore the factory value
after restarting DL-2060. The valid range of “SAVE DATA” is listed in
Table 4-5
.
Table 4-5
No. Items of Save Data
Contents
1
Continuity
Threshold Resistance
2
Diode
Boundary Voltage
3
Temperature (RTD)
Sensor
、
Types
、
Units
、
USER define
、
SPRTD
4
Thermocouple
Types, Units,Simulated
5
Step & Scan
Count, Timer
6
dBm
Reference Resistance
7
MX+B
Values: M, B
8
Reading Hold
Sensitivity Band
9
Input R
DC Input Resistance
10
Interface
USB
、
GPIB ADDR
11
System
Language
、
Beep
How to save the settings
Press MENU and then use
◁
and
▷
to locate “SYSTEM” submenu. Press ENTER to select
it. Again use
◁
and
▷
to locate “INIT MODE” submenu, and then press ENTER to select
it. Use
◁
and
▷
to switch from “SAVE DATA” and “DEFAULT”. Press ENTER on “SAVE
DATA” to save your current setting, or press ENTER on “DEFAULT” to return to default
setting. The locations of the buttons are shown with red rectangle frames in
Figure
4-33
.
Figure 4-33
Summary of Contents for DL-2060