51
annunciator on the display indicates the enabling of the external hardware trigger.
The folloeing figure shows level and timming of a trigger pules.
Figure EXT TRIG (Input) pulse
T
riggere edge
TTL: H
(5V)
TTL: L
(0V) > 1µs
VM COMP (Output)
The rear-panel VM COMP (Voltmeter complete) terminal provides a low-true pulse (TTL
level) after the completion of each measurement. VM COMP and external trigger (see
below) implement a standard hardware handshake sequence between measurement
and switching devices.
Figure VM COMP (Output) pulse
T
riggere edge
TTL: H
(5V)
TTL: L
(0V) Approximately 2µs
Remote Interface Operation (By software or internal trigger)
By software trigger:
The software trigger is similar to the single triggering, but instead
of using the “SINGLE” button on the front panel, you send a command from your PC to
the multimeter to generate an event. For using software trigger, use the following
command from your PC terminal to set trigger source:
TRIGger:SOURce BUS
By internal trigger:
The internal trigger is the default trigger mode for the Remote
Interface Operation. In the internal trigger mode, a trigger signal will be issued
whenever the multimeter is in the wait-for-trigger state. To set the internal trigger, use
the following command from your PC terminal:
TRIGger:SOURce IMMediate