124
SYSTem:VERSion?
Query the present SCPI version.
L0
Set to the default identification string.
L1
Set to the compatible identification string.
DATA:POINts?
Query the number of readings stored in the multimeter’s internal memory.
*RST
Reset the multimeter to the power-on configuration. This command can not clear the error queue.
*IDN?
Read the multimeter’s identification string (be sure to dimension a string variable with at least 35
characters).
Other Interface Commands
SYSTem:LOCal
Place the multimeter in the
local
mode. All buttons on the front panel are fully functional.
SYSTem:REMote
Place the multimeter in the
remote
mode. All buttons on the front panel, except the LOCAL button,
are disabled.
B.9 Status Reporting Commands
SYSTem:ERRor?
Query the multimeter’s error queue. Up to 20 errors can be stored in the queue. Errors are
retrieved in first-in-first out (FIFO) order. Each error string may contain up to 80 characters.
STATus:QUEStionable:ENABle <enable value>
Enable bits in the Questionable Data enable register. The selected bits are then
reported to the Status Byte.
STATus:QUEStionable:ENABle?
Query the Questionable Data enable register. The multimeter returns a binary-weighted decimal
representing the bits set in the enable register.