67
3.4 Other System Related Operations
Each system related operation performs a task that is not measurement related but
plays an important role in making your measurements.
3.4.1 Display
DL-2060 has a 5x7 dot matrix, dual-display with three-color
(
White, Red and Yellow
)
annunciators for a better view. A maximum 13 characters are allowed for upper row
dot-matrix display and a maximum 16 characters are allowed for lower row dot-matrix
display
as shown on
Figure 4-25
.
User has an option to turn off the display for a faster
measurement without waiting for display, or when the measurement is done through the
remote interface operations on their PC terminal.
Figure 4-25
When the display is turned off, an “OFF” will be lit at the lower right corner of the display
screen. This doesn't mean the display is POWER-OFF, but only that the measurement
readings will not be sent to the display. This option will accelerate the measurement
process because there is no I/O delay. But turning off the display will not affect the
messages displayed from the RECALL, MENU and CONFIG operations.
User can send a message through the remote interface from their PC terminal to the
lower row display and will substitute the original display.
Default
The display default is “ON”. The “On/Off” selection made by user is stored in a volatile
memory and the default “ON” will be restored when the meter is power-off.
How to control the display
You can control the display through the front panel operation or through the remote
interface operation. The remote interface operation has higher priority over the front
panel operation.
Front Panel Operation
Use the following steps to control the display: Press MENU button then use
◁
and
▷
buttons to locate “SYSTEM” submenu. Press
ENTER to select it, and then again use
◁
and
▷
buttons to locate “DISPLAY” submenu. Press
ENTER to select it, and then find ON