1.1.3.3 To change the current input fuse
The current input terminal is protected respectively by the fuse(3A,250V Fast
Acting, HBC
:
TEXIO part number F50-0268-08) at rear and front panel. Also, it
is equipped by a 7A(7A,250V Fast Acting
:
TEXIO part number F50-0267-08)
fuse in series which will be the double protection of the current measuring
circuit.
Figure 2-15
Warning! Before replacing the current input fuse, make sure the
multimeter is disconnected from the AC power. You must be qualified
personnel to perform this action.
Warning! For continued protection against fire or instrument damage,
only replace fuse with the same type and rating. If the
instrument repeatedly blows fuses, locate and correct the cause
of the trouble before replacing the fuse.
Current Input Fuse Replacement
(3A on the Front & Rear Panels)
Verify that the power-line fuse is good. Replace the fuse if it is damaged. Use only the
same type and same rating fuse.
(3A,250V Fast Acting, HBC
:
TEXIO part number
F50-0268-08) Please follow the steps below to change the fuse.
【
Step 1
】
Verify that the meter is disconnected as shown in
Figure 2-16
.
Figure 2-16