61
How to make a limit testing
The locations of the buttons are shown with red rectangle frames in
Figure 4-19
.
Select a measurement function for purpose except continuity and diode.
Enable the limit operation by pressing SHIFT + LIMITS(RATIO) buttons.
After enabling the limit function, user can set the limit value as mentioned above.
Observe the displayed reading.
※
Note: Press SHIFT+LIMITS(RATIO) again to disable this feature. The “MATH”
anunnciator on the display indicates the state of a mathematical feature.
Figure 4-19
Remote Interface Operation
Use the following commands to enable the limits function or to set the limits:
CALCulate:FUNCtion LIMit
CALCulate:STATe {OFF|ON}
CALCulate:STATe?
CALCulate:LIMit:LOWer {<
value
>|MINimum|MAXimum}
CALCulate:LIMit:LOWer? [MINimum|MAXimum]
CALCulate:LIMit:UPPer {<
value
>|MINimum|MAXimum}
CALCulate:LIMit:UPPer? [MINimum|MAXimum]
Acquisition of limit test's result
The result for Pass/Fail test may be aquired from the USB connector located at rear
panel for DL-2060.
Please refer "4.1 Pass/Fail output from USB connector" for detail.
Summary of Contents for DL-2060
Page 37: ...31 Figure 3 17 1 2 3 5 5 4...
Page 136: ...130...
Page 137: ...131...
Page 138: ...132 D DL 2060 s Dimension...