29
③
Configure the thermocouple type and unit using SHIFT + TCOUPL(TEMP),
such as K TYPE and ºC. When ready, press ENTER button.
④
Press SHIFT + TEMP buttons.
⑤
Take the reading on the display.
Figure 3-12
2.7.2 RTD Measurements
There are three kinds of temperature measurements with RTDs: 2-wire, 3-wire and
4-wire measurements. You will find connection instructions and measuring procedures in
the following sections. The source current for the RTD measurement is 1 mA.
2.7.2.1 2-Wire RTD Measurements
How to measure temperature with 2-Wire RTD
The following
Figure 3-13
shows theory diagram of 2-Wire RTD measurement.
①
Use terminals switch to select front terminals.
②
Insert a specified adapter into the front terminals.
③
Configure sensor type and unit using TEMP and
◁
or
▷
. When ready, press
ENTER button.
④
Press TEMP button.
⑤
Place RTD in the desired position and take the reading on the display.
Figure 3-13